Advertisement

Parameter
Default
Query
Description
Example
RSTATE
Purpose
Format
Parameter
Default
Description
Example
SETACV
Purpose
Format
Parameter
<mask>
Decimal weight
Bit number
4
2
8
3
32
5
64
6
128
7
0
Returns the weighted sum of all enabled status register bits.
RQS?
The RQS command enables one or more bits in the status register. When an enabled condition occurs,
the SRQ line is asserted. All other conditions and bits are not supported, and thus are always 0. To
enable multiple bits, add up the individual bit weights.
The SRQ command must be sent to enable SRQ.
SRQ on error.
RQS 32
SRQ on error and data available.
RQS 160
To recall an instrument configuration from memory.
RSTATE <state>
<state>
Memory option
No extended memory
2002-MEM1 installed
2002-MEM2 installed
SAV0
The RSTATE command recalls a stored state from memory and configures the multimeter for that
state. These states are saved by the SSTATE command. Note that the number of valid saved states
depends on the amount of installed memory as indicated above.
Recall state 0.
RSTATE SAV0
To select the RMS conversion technique used for AC and AC+DC measurements.
SETACV <type>
<type>
Enabled condition
SRQ command executed
Power-on SRQ
Error (consult the error register)
Service Requested (cannot disable this bit)
Data available
Valid <state> parameters
SAV0
SAV0-SAV4
SAV0-SAV9
HP3458A Emulation Mode
H-17

Advertisement

Table of Contents
loading

Table of Contents