Internal Scanning Example With Timer And Delay Options - Keithley 2000 User Manual

Hide thumbs Also See for 2000:
Table of Contents

Advertisement

Figure 3-15
Internal scanning
example with tim-
er and delay op-
tions
SHIFT-CONFIG
TYPE:INT
Note: "Factory setup" on the
MIN CHAN: 1
Model 2000 is assumed.
MAX CHAN: 10
OFF
TIMER?
ON
TIMER? ON
00H:00M:05.000S
RDG CNT: 0010
SCAN
10 channel closures
1 output trigger
STEP
10 channel closures
at 5-second intervals
10 output triggers
RECALL
10 readings
Measurement Options
RDG CNT: 0010
DELAY: MAN
00H:00M:05.000S
STEP
10 channel closures
at 5-second intervals
10 output triggers
SCAN
10 channel closures
at 5-second intervals
1 output trigger
RECALLL
10 readings
3-29