Trigger Subsystem; Initiate Commands; Abort Command - Keithley 2000 User Manual

Hide thumbs Also See for 2000:
Table of Contents

Advertisement

5-70
SCPI Command Reference

Trigger subsystem

:INITiate commands

[:IMMediate]
:INITiate[:IMMediate]
Description
:CONTinuous <b>
:INITiate:CONTinuous <b>
Parameters
Query
Description

:ABORt command

:ABORt Abort operation
Description
The Trigger subsystem is made up of a series of commands and subsystems
to configure the Trigger Model. These commands and subsystems are sum-
marized in Table 5-10.
This command takes the Model 2000 out of the idle state. After all pro-
grammed operations are completed, the instrument returns to the idle state if
continuous initiation is disabled; (see next command).
<b> = 0 or OFF Disable continuous initiation
1 or ON
Enable continuous initiation
:CONTinuous?
Query continuous initiation
When continuous initiation is selected (ON), the instrument is taken out of
the idle state. At the conclusion of all programmed operations, the instru-
ment returns to the top of the trigger model.
NOTE:
With continuous initiation enabled (ON), you cannot use the
:READ? command or set sample count greater than one (see
:SAMPle:COUNt).
When this action command is sent, the Model 2000 aborts operation and re-
turns to the top of the Trigger Model. If continuous initiation is disabled, the
instrument goes to the idle state. If continuous initiation is enabled, opera-
tion continues at the top of the trigger model.
The abort command resets the scan pointer back to the first channel in the
scan list.
Take 2000 out of idle state
Control continuous initiation