General Information

Introduction
This section contains general information about the Model 2000 Multimeter. The information
is organized as follows:
If you have any questions after reviewing this information, please contact your local
Keithley representative or call one of our Applications Engineers at 1-800-348-3735 (U.S.
and Canada only). Worldwide phone numbers are listed at the front of this manual.

Feature overview

The Model 2000 is a 6 -digit high-performance digital multimeter. It has 0.002% 90-day
basic DC voltage accuracy and 0.008% 90-day basic resistance accuracy. At 6
timeter delivers 50 triggered readings/sec over the IEEE-488 bus. At 4
2000 readings/sec into its internal buffer. The Model 2000 has broad measurement ranges:
Some additional capabilities of the Model 2000 include:
Feature overview
Warranty information
Manual addenda
Safety symbols and terms
Specifications
Inspection
Options and accessories
DC voltage from 0.1 V to 1000V.
AC (RMS) voltage from 0.1 V to 750V, 1000V peak.
DC current from 10nA to 3A.
AC (RMS) current from 1 A to 3A.
Two and four-wire resistance from 100µ to 120M .
Frequency from 3Hz to 500kHz.
Thermocouple temperature from -200°C to +1372°C.
Full range of functions — In addition to those listed above, the Model 2000 functions
include period, dB, dBm, continuity, diode testing, mX+b, and percent.
Optional scanning — For internal scanning, options include the Model 2000-SCAN, a
10-channel, general-purpose card, and the Model 2001-TCSCAN, a 9-channel, thermo-
couple card with a built-in cold junction. For external scanning, the Model 2000 is com-
patible with Keithley's Model 7001 and 7002 switch matrices and cards.
Programming languages and remote interfaces — The Model 2000 offers three program-
ming language choices (SCPI, Keithley Models 196/199, and Fluke 8840A/8842A) and
two remote interface ports (IEEE-488/GPIB and RS-232C).
Reading and setup storage — Up to 1024 readings and two setups (user and factory de-
faults) can be stored and recalled.
Closed-cover calibration — The instrument can be calibrated either from the front panel
or remote interface.
digits, the mul-