:TIMer <n>
:TRIGger:[SEQuence[1]]:TIMer <n>
Parameters
<n> = 0.001 to 999999.999
Query
:TIMer?
Description
These commands are used to set the interval for the timer. Note that the timer
is in effect only if the timer is the selected control source.
:SIGNal
:TRIGger[:SEQuence[1]]:SIGNal
Description
This action command is used to bypass the specified control source when you
do not wish to wait for the programmed event. Keep in mind that the instru-
ment must be waiting for the appropriate event when the command is sent.
Otherwise, an error occurs and this command is ignored.
:SAMPle Command
:SAMPle:COUNt <NRf>
Parameter
<NRf> = 1 to 1024
Query
:COUNt?
Description
This command specifies the sample count. The sample count defines how
many times operation loops around in the trigger model to perform a device
action.
NOTE
Specify timer interval in seconds
Query programmed timer interval
Query the sample count
If sample count is >1, you cannot use the :READ? command if
there are readings stored in the buffer.
SCPI Command Reference
Set interval for measure layer timer
Bypass measure control source
Set sample count