:REAL:OFFSet <n>
[:SENSe[1]]:TEMPerature:TCouple:RJUNction[1]:REAL:OFFSET <n>
Parameters
Query
Description

:DIODe command

:RANGe[:UPPer] <NRf>
[:SENSe[1]]:DIODe:CURRent:RANGe[:UPPer] <NRf>
Parameters
Query
Description

:CONTinuity command

:THReshold <n>
[SENSe[1]]:CONTinuity:THReshold <NRf>
Parameters
Query
<n> = -0.09999 to 0.09999
DEFault
MINimum
MAXimum
:OFFSet?
Query voltage offset
:OFFSet? DEFault
Query *RST default voltage offset
:OFFSet? MINimum
Query lowest allowable voltage offset
:OFFSet? MAXimum Query largest allowable voltage offset
This command is used to specify the offset voltage at 0°C for the specified
reference junction.
<NRf> = 0 to 1e-3 Specify diode test current
[UPPer]?
Query selected range
There are three current ranges available for the diode test: 10µA range,
100µA range and the 1mA range. Range is selected by using this command
to specify the expected current for the diode under test. The instrument will
then automatically select the appropriate range.
<NRf> = 1 to 1000 Specify threshold in ohms
:THReshold?
Query threshold resistance
This command is used to specify the threshold resistance for the continuity
test. Continuity occurs when the measurement is less than or equal to the
specified threshold level.
SCPI Command Reference
Specify voltage offset at 0°C
0.05463
-0.09999
0.09999
Select current range for diode test
Specify threshold resistance
5-51