Keithley 2000 User Manual page 133

Hide thumbs Also See for 2000:
Table of Contents

Advertisement

Program Fragment
PRINT #1, "output 16; :syst:pres"
PRINT #1, "output 16; :init:cont off;:abort"
PRINT #1, "output 16; :trig:coun 1; sour tim"
PRINT #1, "output 16; :samp:coun 5"
PRINT #1, "output 16; :init; *opc?"
PRINT #1, "enter 16"
LINE INPUT #2, a$
PRINT a$
*OPT? — Option Identification Query
Description
The response message indicates the presence or absence of an optional scanner card. For ex-
ample:
0
200X-SCAN
*RCL — Recall
Parameters
<NRf>=0
Description
Use this command to return the Model 2000 to the configuration stored in memory. The *SAV
command is used to store the setup configuration in memory location.
Only one setup configuration can be saved and recalled.
The Model 2000 ships from the factory with :SYSTen:PRESet defaults loaded into the avail-
able setup memory. If a recall error occurs, the setup memory defaults to the :SYSTem:PRESet
values.
No scanner card installed
Scanner card installed
Return to setup stored in memory
Remote Operation
'Select defaults
'Place 2000 in idle
'Program for 5 measurements and
stop (idle)
'Start measurements and send
*opc?
'Get response when 2000 goes into
idle
'Read contents of Output Queue
'Display the ASCII "1"
Determine if an option is installed
4-47