:TRIGger commands

:COUNt <n>
:TRIGger[:SEQuence[1]]:COUNt <n>
Parameters
Query
Description
<n> = 1 to 9999
Specify count
INF
Sets count to infinite
DEFault
Sets count to 1
MINimum
Sets count to 1
MAXimum
Sets count to 9999
:COUNt?
:COUNt? DEFault
:COUNt? MINimum
:COUNt? MAXimum
This command is used to specify how many times operation loops around in
the trigger operation. For example, if the count is set to 10, operation con-
tinues to loop around until 10 device actions are performed. After the 10th
action, operation proceeds back up to the start of the trigger model. Note that
each loop places operation at the control source where it waits for the pro-
grammed event.
SCPI Command Reference
Set measure count
Queries programmed count
Queries *RST default count
Queries lowest allowable count
Queries largest allowable count