Table D-1 (cont.)

Models 196/199 device-dependent command summary

Mode
Rate
Trigger mode
Reading mode
Data store size
Interval
Value
V±nn.nnnn or
V±n.nnnnnnE+n
Default conditions
Data format
SRQ
EOI and bus hold-off

Models 196/199 and 8840A/8842A Commands

Command
Description
S0
0.1 PLC integration
S1
Line cycle integration (16.67msec, 60Hz; 20msec,
50Hz)
S2
10 PLC (166.67msec integration, 60Hz;
200msec integration, 50Hz)
T0
Continuous on Talk
T1
One-shot on Talk
T2
Continuous on GET
T3
One-shot on GET
T4
Continuous on X
T5
One-shot on X
T6
Continuous on External Trigger
T7
One-shot on External Trigger
B0
Readings from A/D converter
B1
Individual readings from data store
B2
All readings from data store (buffer dump)
I0
Disable data store
In
Data store of n (n=1 to 500), fill and stop
Q0
Default interval, 175msec (SELECT OFF)
Qn
n=interval in milliseconds (15msec to 999999msec)
Zero value, simulated reference junction temperature
L0
Restore factory default conditions and save (L1)
L1
Save present machine states as default conditions
G0
Reading with prefix.
G1
Reading without prefix.
G2
Reading and buffer location with prefix.
G3
Reading and buffer location without prefix.
G4
Reading and channel with prefix.
G5
Reading and channel without prefix.
G6
Reading, buffer location, and channel with prefix.
G7
Reading, buffer location, and channel without prefix.
M0
Disable
M1
Reading overflow
M2
Data store full
M4
Data store half full
M8
Reading done
M16
Ready
M32
Error
K0
Enable EOI and bus hold-off on X
K1
Disable EOI, enable bus hold-off on X
K2
Enable EOI, disable bus hold-off on X
K3
Disable both EOI and bus hold-off on X