:COUNt <n>
:TRIGger[:SEQuence[1]]:COUNt <n>
Parameters
Query
Description
:DELay <n>
:TRIGger[:SEQuence[1]]:DELay <n>
Parameters
Query
Description
<n> = 1 to 9999
Specify count
INF
Sets count to infinite
DEFault
Sets count to 1
MINimum
Sets count to 1
MAXimum
Sets count to 9999
:COUNt?
:COUNt? DEFault
:COUNt? MINimum
:COUNt? MAXimum
This command is used to specify how many times operation loops around in
the trigger operation. For example, if the count is set to ten, operation
continues to loop around until ten device actions are performed. After the
tenth action, operation proceeds back up to the start of the trigger model.
Note that each loop places operation at the control source where it waits for
the programmed event.
<n> = 0 to 999999.999
DEFault
MINimum
MAXimum
:DELay?
:DELay? DEFault
:DELay? MINimum
:DELay? MAXimum
The delay is used to delay operation of the trigger model. After the
programmed event occurs, the instrument waits until the delay period expires
before performing the Device Action in the Trigger Model.
The delay time can also be set by using the AUTO parameter. If AUTO is set
to 1 or on, the delay period is enabled and will occur. If AUTO is set to 0 or
off, the delay period is not enabled, and no delay will occur.
Set measure count
Queries programmed count
Queries *RST default count
Queries lowest allowable count
Queries highest allowable count
Set trigger model delay
Specify delay in seconds
0 second delay
0 second delay
999999.999 second delay
Query the programmed delay
Query the *RST default delay
Query the lowest allowable delay
Query the highest allowable delay