Simulation - Endress+Hauser Proline Prosonic Flow W 400 Operating Instructions Manual

Ultrasonic time-of-flight flowmeter
Hide thumbs Also See for Proline Prosonic Flow W 400:
Table of Contents

Advertisement

Commissioning
Parameter
Device reset
*
Visibility depends on order options or device settings
114
Parameter overview with brief description
Description
Reset the device configuration - either entirely or in part - to a
defined state.
10.6

Simulation

Via the Simulation submenu, it is possible to simulate various process variables in the
process and the device alarm mode and verify downstream signal chains (switching valves
or closed-control loops). The simulation can be performed without a real measurement (no
flow of medium through the device).
The parameters displayed depend on:
• The selected device order
• The set operating mode of the pulse/frequency/switch outputs
Navigation
"Diagnostics" menu → Simulation
Simulation
Assign simulation process variable
Process variable value
Status input 1 simulation
Input signal level 1
Current output 1 simulation
Current output value
Frequency output 1 to n simulation
Frequency output 1 to n value
Pulse output simulation 1 to n
Pulse value 1 to n
Switch output simulation 1 to n
Switch state 1 to n
Device alarm simulation
Proline Prosonic Flow W 400 HART
Selection
• Cancel
• To delivery settings
• Restart device
*
• Restore S-DAT backup
→  115
→  115
→  115
→  115
→  115
→  115
→  115
→  115
→  115
→  115
→  115
→  115
→  115
Endress+Hauser

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Proline Prosonic Flow W 400 and is the answer not in the manual?

Table of Contents