Commissioning
Parameter
Measuring point configuration
Medium
Medium temperature
Sound velocity calculation mode
Sound velocity
Viscosity
Pipe material
Pipe sound velocity
Pipe dimensions
88
Parameter overview with brief description
Prerequisite
–
–
–
The Liquid hydrocarbons
option is selected in Medium
parameter.
The User-specific liquid
option is selected in Medium
parameter.
The User-specific liquid
option is selected in Medium
parameter.
–
The Unknown pipe material
option is selected in the Pipe
material parameter.
–
Description
Select configuration for the
measuring point.
Select the medium.
Enter the medium temperature
for the installation.
Select the process variable to
use to calculate the sound
velocity at installation.
Enter the medium' s sound
velocity for the installation.
Enter medium viscosity at
installation temperature.
Select pipe material.
Enter sound velocity of pipe
material.
Select if pipe dimensions are
defined by diameter or
circumference.
Proline Prosonic Flow W 400 HART
Selection / User
Factory setting
entry / User
interface
• 1 measuring point
Depending on the
- signal path 1
sensor version
• 1 measuring point
*
- signal path 2
• 1 measuring point
*
- 2 signal paths
• Water
–
• Sea water
• Distilled water
• Ammonia NH3
• Benzene
• Ethanol
• Glycol
• Milk
• Methanol
• User-specific liquid
–200 to 550 °C
–
• Fixed value
–
• API gravity
• Density
• Reference density
200 to 3 000 m/s
–
0.01 to 10 000 mm²/
–
s
• Carbon steel
–
• Ductile cast iron
• Stainless steel
• 1.4301 (UNS
S30400)
• 1.4401 (UNS
S31600)
• 1.4550 (UNS
S34700)
• Hastelloy C
• PVC
• PE
• LDPE
• HDPE
• GRP
• PVDF
• PA
• PP
• PTFE
• Pyrex glass
• Asbestos cement
• Copper
• Unknown pipe
material
800.0 to 3 800.0 m/s –
• Diameter
–
• Pipe circumference
Endress+Hauser
Need help?
Do you have a question about the Proline Prosonic Flow W 400 and is the answer not in the manual?