Mounting procedure
Measuring pipe material: steel, cast iron
Measuring pipe wall thickness [mm (in)]
1.0 to 1.9 (0.04 to 0.07)
> 1.9 to 2.2 (0.07 to 0.09)
> 2.2 to 2.8 (0.09 to 0.11)
> 2.8 to 3.4 (0.11 to 0.13)
> 3.4 to 4.2 (0.13 to 0.17)
> 4.2 to 5.9 (0.17 to 0.23)
> 5.9 (0.23)
1)
Table shows a typical selection: In critical cases (large pipe diameter, liner, gas or solid inclusions) the optimum sensor type may differ from these
recommendations.
Measuring pipe material: Steel, cast iron with wall thicknesses > 5.9 mm (0.23 in)
Nominal diameter [mm (")]
15 to 50 (¹⁄₂ to 2)
> 50 to 300 (2 to 12)
> 300 to 1 000 (12 to 40)
> 1 000 to 4 000 (40 to 160)
1)
Table shows a typical selection: In critical cases (large pipe diameter, liner, gas or solid inclusions) the optimum sensor type may differ from these
recommendations.
Measuring pipe material: plastic
Nominal diameter [mm (")]
15 to 50 (¹⁄₂ to 2)
> 50 to 80 (2 to 3)
> 80 to 150 (3 to 6)
> 150 to 200 (6 to 8)
> 200 to 300 (8 to 12)
> 300 to 400 (12 to 16)
> 400 to 500 (16 to 20)
> 500 to 1 000 (20 to 40)
> 1 000 to 4 000 (40 to 160)
1)
Table shows a typical selection: In critical cases (large pipe diameter, liner, gas or solid inclusions) the optimum sensor type may differ from these
recommendations.
26
Kinematic viscosity cSt [mm
0 < v ≤ 10
Converter frequency (sensor version/number of traverses)
2 MHz (C-200 / 2)
1 MHz (C-100 / 2)
2 MHz (C-200 / 2)
1 MHz (C-100 / 2)
2 MHz (C-200 / 2)
1 MHz (C-100 / 2)
Selection according to table: "Measuring pipe material: steel, cast iron > 5.9 mm (0.23 in)"
Kinematic viscosity cSt [mm
0 < v ≤ 10
Converter frequency (sensor version/number of traverses)
2 MHz (C-200)
1 MHz (C-100)
Kinematic viscosity cSt [mm
0 < v ≤ 10
Converter frequency (sensor version/number of traverses)
5 MHz (C-500 / 2)
2 MHz (C-200 / 2)
1 MHz (C-100 / 2)
1 MHz (C-100 / 2)
1 MHz (C-100 / 2)
1 MHz (C-100 / 1)
1 MHz (C-100 / 1)
0.3 MHz (C-030 / 1)
0.3 MHz (C-030 / 1)
Proline Prosonic Flow W 400 HART
2
/s]
10 < v ≤ 100
2 MHz (C-200 / 1)
1 MHz (C-100 / 1)
1 MHz (C-100 / 1)
1 MHz (C-100 / 1)
2 MHz (C-200 / 1)
1 MHz (C-100 / 1)
2
/s]
10 < v ≤ 100
5 MHz (C-500)
1 MHz (C-100)
0.3 MHz (C-030)
0.3 MHz (C-030)
2
/s]
10 < v ≤ 100
5 MHz (C-500 / 2)
1 MHz (C-100 / 2)
1 MHz (C-100 / 2)
0.3 MHz (C-030 / 2)
0.3 MHz (C-030 / 2)
0.3 MHz (C-030 / 2)
0.3 MHz (C-030 / 1)
0.3 MHz (C-030 / 1)
100 < v ≤ 1000
1)
2 MHz (C-200 / 1)
1 MHz (C-100 / 1)
1 MHz (C-100 / 1)
1 MHz (C-100 / 1)
1 MHz (C-100 / 1)
0.3 MHz (C-030 / 2)
100 < v ≤ 1000
1)
1 MHz (C-100)
0.3 MHz (C-030)
100 < v ≤ 1000
1)
5 MHz (C-500 / 2)
0.3 MHz (C-030 / 2)
0.3 MHz (C-030 / 2)
0.3 MHz (C-030 / 2)
0.3 MHz (C-030 / 2)
0.3 MHz (C-030 / 1)
0.3 MHz (C-030 / 1)