Error message
Mode change failed
16.2.4 Status signals
Battery charger is connected
LED on
LED on bat
power
tery charger
supply
Battery charger disconnected
Beep
Vibration
signal
signal
1 x short
1 x short
–
3 x short
–
–
Battery charge level
Battery char
ger
Battery is being charged, battery charge level is less than 50%
Cause
An attempt was made to
switch to a different
MyMode while the com
ponent was in motion
(e. g. while walking)
A current connection to
the component was
interrupted
Event
Power supply and battery charger operational
Event
Self-test completed successfully. Product is operational.
Maintenance note
Conduct the self-test again by connecting/disconnecting the battery
charger. If the beep signal sounds again, visit your O&P professional
soon. If necessary, he or she will forward the product to an authorised
Ottobock Service Center.
The product can be used without restrictions. However, vibration signals
may not be generated.
Conduct the self-test again by connecting/disconnecting the battery
charger. If no beep and/or vibration signal is emitted after connect
ing/disconnecting the battery charger again, the product must be
inspected by the O&P professional.
Correction
For safety reasons, switching MyModes is
only permitted when components are at
rest, e. g. while standing or sitting.
Check the following points:
•
Distance from the component to the
device
•
Charge level of the component's bat
tery
•
Bluetooth of the component switched
on? (Switching Bluetooth of the com
ponent on/off)
•
Hold the component with the sole of the
foot facing up to make the component
"visible" for 2 minutes.
•
If multiple components were stored,
was the correct component selected?
41