Parameter
Volume
1
To use these functions in the Cockpit app, they need to be enabled in the adjustment software.
8.2.3 Overview of adjustment parameters in MyModes
The parameters in the MyModes describe the static behaviour of the prosthesis for a specific
motion pattern such as cross-country skiing. Damping behaviour is not automatically controlled
and adjusted in MyModes.
The following parameters can be modified in MyModes:
Parameter
Basic flex.
Gain
Basic ext.
Locking angle
Adjustment
Setting
software
range, app
range
0 to 4
0 to 4
Adjustment
Setting
software
range, app
range
0–200
+/- 20 of the
configured
value
0–100
+/- 10 of the
configured
value
0–60
+/- 20 of the
configured
value
0–90
+/- 10 of the
configured
value
Meaning
Volume of beep signal for confirma­
tion tones (e.g. when checking the
charge level, switching MyModes).
The "0" setting deactivates the aud­
ible feedback signals. However, warn­
ing signals are still generated if errors
occur.
Meaning
Level of flexion resistance when the
knee joint begins to flex
Increase in flexion resistance (starting
with the "Basic flex." parameter)
when flexing the knee joint. The knee
joint locks at a certain flexion angle,
which depends on the settings for the
"Basic flex." and "Gain" parameters.
Level of extension resistance
Angle up to which the knee joint can
be extended.
Information: If this parameter is >0,
the knee joint is locked in a flexed
position in the extension direction. To
unlock it, take all weight off the pros­
thesis and tilt it back for at least
2 seconds. This enables extension of
the joint independently of the settings
for the "Basic ext." and "Locking
angle" parameters. This may be
necessary to switch to basic mode
using a movement pattern.