16.2 Operating states/error signals
The prosthesis indicates operating states and error messages through beeps and vibration sig
nals.
16.2.1 Signals for operating states
Battery charger connected/disconnected
Beep signal
1x short
–
1x short
Mode switching
INFORMATION
If the Volume parameter is set to '0' in the Cockpit app, there are no beep signals (see
page 25).
Beep signal
Vibration
signal
1 x short
1 x short
1 x short
1 x short
1 x short
1 x short
2 x short
2 x short
3 x short
3 x short
16.2.2 Warnings/error signals
Error during use
Beep signal
–
–
38
Vibration signal
–
3x short
1x before beep signal Battery charger disconnected after start of char
Additional action per
formed
Mode switching using the
Cockpit app
Bouncing on the forefoot fol
lowed by holding still for
1 second in the walking pos
ition
Prosthetic leg moved next to
contralateral leg, set down
and kept still for 1 second
Prosthetic leg moved next to
contralateral leg, set down
and kept still for 1 second
Prosthetic leg moved next to
contralateral leg, set down
and kept still for 1 second
Vibration signal
1 x long at interval of
approx. 5 seconds
3 x long
Event
Battery charger is connected or
battery charger already disconnected prior to
start of charging mode
Charging mode started (3 sec. after connecting
battery charger)
ging mode
Result
Mode switching is performed
using the Cockpit app.
Bouncing pattern recognised.
Switching to basic mode
(mode 1) carried out.
Switching to MyMode 1 (mode 2)
carried out.
Switching to MyMode 2 (mode 3)
carried out.
Event
Overheated hydraulic
unit
Battery charge level
under 25%
Required action
Reduce activity.
Charge battery soon.