Otto Bock C-Leg 3C98-3 Instructions For Use Manual page 26

Hide thumbs Also See for C-Leg 3C98-3:
Table of Contents

Advertisement

8.2.1 Changing the prosthesis setting using the cockpit app
8.2.2 Overview of adjustment parameters in basic mode
The parameters in basic mode describe the dynamic behaviour of the prosthesis in a normal gait
cycle. These parameters act as basic settings for automatically adjusting the damping behaviour
to the current motion situation (e.g. ramps, slow walking speed, etc.).
The stance function and/or the sitting function can also be activated/deactivated. Further informa­
tion on the stance function (see page 22). Further information on the sitting function (see
page 23).
The following parameters can be modified:
Parameter
Resistance
1
Stance function
1
Sitting function
Acoustic feedback sig­
nal
26
1) Once the component is connected and in the desired mode,
tap the  icon in the main menu.
→ The navigation menu opens.
2) Tap the "Settings" menu option.
→ A list appears with the parameters for the currently
selected mode.
3) Change the setting of the desired parameter by tapping the
"<", ">" icons.
INFORMATION: The O&P professional's setting is
marked and, after the setting has been changed, can be
restored by tapping the "Standard" button.
Adjustment
Setting
software
range, app
range
120 to 190
+/- 10 of the
configured
value
0/Off – deacti­
vated
1/On – activ­
ated
0/Off – deacti­
vated
1/On – activ­
ated
On/Off
Meaning
Flexion resistance while sitting down,
in the stance phase, while walking on
ramps and stairs.
Information about this function is
provided in the section "Stance func­
tion" (see page 22)
When the function is activated, the
resistance in the flexion direction
while sitting is reduced in addition to
the reduction of resistance in the
extension direction.
Acoustic feedback for switching
between the stance and swing phase.