Otto Bock C-Leg 3C98-3 Instructions For Use Manual page 39

Hide thumbs Also See for C-Leg 3C98-3:
Table of Contents

Advertisement

Beep signal
10 x long
30 x long
Vibration signal
5 x long
10 x long
1 x long, 1 x short
repeated every
3 seconds
Continuous
Event
Charge level under
15%
Charge level 0%
After the beep and
vibration signals, the
product switches to
empty battery mode
and then switches
off.
Severe error/indic­
ation of safety
mode activation
e.g. one or more
sensors are not oper­
ational.
Total failure
Electronic control no
longer possible.
Safety mode active or
undetermined valve
state. Undetermined
product behaviour.
Required action
Charge battery immedi­
ately; the product will be
switched off after the next
warning signal.
Charge the battery.
Walking possible with
restrictions. Please note
the possible change in
flexion/extension resist­
ance.
Attempt to reset this error
by connecting/discon­
necting the battery char­
ger. The battery charger
must be connected for at
least 5 seconds before it
is disconnected.
If the error persists, use
of the product is prohib­
ited. The product must be
inspected immediately by
an O&P professional.
Attempt to reset this error
by connecting/discon­
necting the battery char­
ger.
If the error persists, use
of the product is prohib­
ited. The product must be
inspected immediately by
an O&P professional.
39