4) Following the beep and vibration signal, move the prosthetic leg next to the contralateral leg,
set it down and keep still for about 1 second.
→ A confirmation signal will sound to indicate that the prosthesis has successfully switched to
the corresponding MyMode (2 times = MyMode 1, 3 times = MyMode 2).
INFORMATION: If this confirmation signal does not sound, the leg with the prosthesis
was not correctly repositioned and kept still. Repeat the process to correctly switch to
the required mode.
9.3 Switching from a MyMode back to basic mode
Information on switching
Regardless of the configuration of additional MyModes in the adjustment software, it is always
possible to switch back to basic mode (mode 1) with a motion pattern.
It is always possible to switch back to basic mode (mode 1) by connecting/disconnecting the
battery charger.
Before the first step, always check whether the selected mode corresponds to the required
motion type.
If the Volume parameter is set to "0" in the Cockpit app, there are no beep signals (see
page 25).
Switching process
3x – 5x
ð
1) Position the prosthetic leg back slightly.
2) While maintaining constant contact with the floor, bounce on the forefoot at least 3 times but
not more than 5 times.
3) Keep the prosthetic leg still in this position (lunge position) for about 1 second without lifting
the leg. Taking the weight off is no longer necessary.
→ A beep and vibration signal will occur to confirm that the movement pattern has been
recognised.
INFORMATION: If this beep and vibration signal is not emitted, the requirements
were not met while bouncing.
4) Move the prosthetic leg in next to the contralateral leg, set it down and keep it still for approx.
1 second.
→ A confirmation signal will sound to indicate that the prosthesis has successfully switched over
to basic mode.
INFORMATION: If this confirmation signal does not sound, the leg with the prosthesis
was not correctly repositioned and kept still. Repeat the process to correctly switch to
the required mode.

10 Additional operating states (modes)

10.1 Empty battery mode
Beeps and vibration signals are emitted if the available battery charge level is 0% (see page 38).
During this time, damping settings are set to their safety mode values. The prosthesis is then
switched off. You can switch back to basic mode (mode 1) from empty battery mode by charging
the product.
1s
1x
+
ð
1x
1s
ð
ð
1x
1x
Basic mode
+
ð
þ
31