Switching From A Mymode Back To Basic Mode; Additional Operating States (Modes); Empty Battery Mode - Otto Bock X2 Series Instructions For Use Manual

Bionic prosthetic system
Hide thumbs Also See for X2 Series:
Table of Contents

Advertisement

Available languages

Available languages

10.4 Switching from a MyMode back to basic mode

Information on switching
Regardless of the configuration of additional MyModes in the adjustment software, it is always
possible to switch back to basic mode (mode 1) with a motion pattern.
It is always possible to switch back to basic mode (mode 1) by connecting/disconnecting the
battery charger.
Before the first step, always check whether the selected mode corresponds to the required
motion type.
Requirements for successful switching using motion patterns
The following points must be observed to carry out switching successfully:
Position the prosthetic leg back slightly and bounce on the forefoot with the leg extended
while maintaining constant contact with the floor.
Weight must be placed on the forefoot while bouncing.
The load may not be taken off the foot fully when relieving the load.
Switching process
≥3x
ð
1) Position the prosthetic leg back slightly.
2) While maintaining constant contact with the floor and with the leg extended, bounce on the
forefoot at 3 times or more.
3) Keep the prosthetic leg still in this position with no load.
→ A beep and vibration signal will sound to confirm that the movement pattern has been
recognised.
INFORMATION: If this beep and vibration signal does not sound, the requirements
were not observed when bouncing the foot or mute mode (silent mode) is activ­
ated. For more information about mute mode, see the section "Mute mode" (see
Page 87).
4) After the beep and vibration signal sounds, keep the prosthetic leg extended and still for 1
second.
→ A confirmation signal will sound to indicate that the prosthesis has successfully switched over
to basic mode.
INFORMATION: If this confirmation signal does not sound, the leg with the prosthesis
was not correctly kept still or mute mode (silent mode) is activated. Repeat the pro­
cess to correctly switch to the required mode. For more information about mute mode,
see the section "Mute mode" (see Page 87).

11 Additional operating states (modes)

11.1 Empty battery mode

Beeps and vibration signals are emitted if the available battery charge level is 5% (see Page 99).
During this time, damping settings are set to their safety mode values. This may be low or high
depending on the setting in the adjustment software. The prosthesis is then switched off. You can
switch back to basic mode (mode 1) from empty battery mode by charging the product.
1x
ð
+
1s
1x
1x
ð ð
1x
ð
þ
þ
Basic mode
+
91

Advertisement

Chapters

Table of Contents
loading

This manual is also suitable for:

3b5-x23b5-x2-st

Table of Contents