Upgrade kit instructions
00825-0200-3810
12. Insert tubing over wires, inner insulation and under shield approximately 1.5-in. of
13. Insert and secure transducer cables through cable glands verify Line Mark is as
14. Using a torque screwdriver set to 78±7 oz.-in. secure the wires to the terminal plugs
15. Verify the cable lengths are still an appropriate length and if necessary, cut the
16. Ensure that the contacts clamp on the bare wires and not on the wire insulation
Emerson.com
Figure 2-8: Transducer Cable Cut Instructions
shield should overlap.
referenced in
Figure
2-7.
Figure 2-9: Line Mark
of the Acquisition Module.
transducer cables to the correct length with allowance for the wiring terminations
and connector placement.
when the wires are being terminated on the connector.
a) Leave the connector plugged into the Acquisition board while terminating the
individual wires.
b) Tighten the connector screws on the either end of the connector after
terminating all the wires on the applicable connector.
c) Repeat this procedure for the transducer cables on the other side of the
meter.
Install
February 2024
19