Operate
February 2024
36
Figure 3-4: Edit/Compare Configuration Dialog
a) Select the final 3800 Series configuration file created in
before removing the 3800 Series Electronics
Name, 3800 Series Final Config, 3-4-2005 10-15-02 AM.cfg").
b) Click Open. The file opens and you returned to the
Configuration
Figure 3-5: Open Configuration File Dialog
c) Click Convert (see
the meter and modifies it with the 3800 configuration.
dialog.
Figure
3-6). MeterLink
reads the 3810 configuration from
Upgrade kit instructions
00825-0200-3810
Step 2
of
Procedure
of this document (i.e., "Meter
Edit/Compare
3810 Series Electronics