Upgrade kit instructions
00825-0200-3810
Emerson.com
Figure 3-11: Field Setup Wizard Frequency and Digital Output Sources Page
The meter has user-configurable selections for either a frequency output or digital
status (FODO). Use the pull-down menus to edit these parameters.
(3) Frequency/Digital Outputs:
FODO1 (four possible output configurations)
FODO2 (eight possible output configurations)
FODO3 (eight possible output configurations)
Frequency or Digital Outputs (FODO 1) source selections:
FO1A, DO1A, FO1B, DO1B
Frequency or Digital Outputs (FODO 2) source selections:
FO1A, DO1A, FO1B, DO1B, FO2A, DO2A, FO2B, DO2B
Frequency or Digital Outputs (FODO 3) source selections:
FO1A, DO1A, FO1B, DO1B, FO2A, DO2A, FO2B, DO2B
Mode options:
Open Collector (requires external excitation supply voltage and pull-up resistor)
TTL (internally powered by the meter 0-5 VDC signal)
Channel B Phase options:
Lag forward, Lead reverse (Phase B lags Phase A while reporting forward flow,
leads Phase A while reporting reverse flow)
Lead forward, Lag reverse (Phase B leads Phase A while reporting forward flow,
lags Phase A while reporting reverse flow)
Operate
February 2024
41