Install
February 2024
27. Install a ground wire to the ground lug on the side of the upper enclosure.
28. Attach flex conduit to the conduit ports on the transmitter electronics enclosure.
29. Apply the sealing compound to the conduit seal fittings and allow to set in
30. Apply electrical power to the system and verify the field connections are working
31. Do not install end caps until you configure the meter.
28
in the electronics. Refer to the System Wiring Drawing in
drawings.
Figure 2-17: Transmitter Electronics Enclosure Ground Wire
A. Attach ground wire to external ground lug
Before applying power to the meter, ensure the atmosphere is proven safe using an
intrinsically safe gas detector.
 WARNING
EXPLOSION HAZARD
Do not apply power to the meter with the end caps removed unless you are in a
non-hazardous or explosive-free environment.
Failure to comply could result in death, serious injury and possible equipment
damage.
accordance with manufacturer specifications.
correctly.
This completes the installation of the hardware associated with the 3810 Series
Electronics meter upgrade procedure.
Upgrade kit instructions
00825-0200-3810
3810 Series engineering
3810 Series Electronics