Plan
February 2024
7. Loosen the cable glands on the base enclosure and pull all of the transducer cables
8. Unscrew the three screws holding the Acquisition module in the base enclosure and
9. Remove the RTV with a flat blade screwdriver or utility knife around the perimeter of
10. Remove the base enclosure from the meter body.
11. If RTV silicone adhesive/sealant was used to seal the base enclosure to the meter
14
b) Disconnect the connectors and remove each of the transducer wires from the
connector.
out of the base enclosure.
remove the Acquisition module.
The four mounting bolts holding the enclosure cover to the meter body are sealed
with an RTV silicone adhesive/sealant.
the recessed opening and the RTV eases the removal of the sealant.
a) Use the 7/16-in. socket wrench to remove the four mounting bolts.
Figure 1-10: Remove Old RTV
body, remove the adhesive/sealant so the 3810 base enclosure will have a clean
mounting surface.
The 3800 Series electronics should be completely removed from the meter body at
this time.
Upgrade kit instructions
00825-0200-3810
3810 Series Electronics