Operate
February 2024
3. Use the
38
Figure 3-8: Converted Configuration with Highlighted Changes
d) Click Write Checked to write the converted configuration to the meter (see
Figure
3-8). If an error occurs, MeterLink
the nature of the problem. Manually edit the parameter(s) with errors to
successfully complete the download.
e) Click Write Checked again after modifying the parameter(s). Repeat this until
the write occurs with no errors.
f) Save the 3810 Series electronics configuration file to your PC.
g) With the converted configuration still displayed, from the Edit/Compare
Configuration dialog, click the Meter radio button, located next to the
Compare button (see
h) Compare the file to verify that this saved configuration matches the meter
configuration. If an error occurs, go back to the beginning of the
try again until no errors are reported.
Meter|Monitor
the meter is operating correctly.
will display a message indicating
Figure
3-8).
menu path to display the Monitor window and verify that
Upgrade kit instructions
00825-0200-3810
Step 2
and
3810 Series Electronics