18
|
The powered footrest
Voltages Elegance/Alliance
E16-3
24 V=
E16-1
0 V=
E15-3
10 V=
16 V=
E15-4
0 V=
E15-2
24 V=
E13-1/2
0 V=
13 V=
18V=
A powered footrest requires PCB version Q or higher.
The powered footrest doesn't function
The Voltage on E15-3 (Elegance) or C18-3 (Classic) is 0V=
•
Wire break between E1-10 and E15-3 (C1-10 and C18-3)
•
The emergency stop is activated
•
Wire break between E11-3 and E11-4 (C14-3 and C14-4)
•
The key switch is off
•
Wire break between E9-2 an d E9-3 (C12-2 and C12-3)
•
No power from the batteries
Interruptions in the chair position switch and the final limit switch and interruptions in the direction
safety lines will not affect the functioning of the powered footrest
The powered footrest doesn't function
The voltage on E15-3 (Elegance) or C18-3 (Classic) is 10V= or 16 V=
•
Wire break between the batteries and E16-3 (C19-3)
•
Wire break between E13-1 and E13-2 (C16-1 and C16-2), (The light on the Power opt printed
circuit illuminates green for a few seconds)
•
Powered footrest control is faulty
The Lighting in the powered footrest does not illuminate; the powered footrest does function
•
The LED lighting is faulty
•
Wire break between E14-2 and E15-4 (C17-2 and C18-4)
always
C19-3
always
C19-1
up
C18-3
down
always
C18-4
down
C17-2
no movement
C16-1/2
down
up
Technical manual
Voltages Classic
24 V=
always
0 V=
always
10 V=
up
16 V=
down
0 V=
always
24 V=
down
0 V=
no movement
13 V=
down
18V=
up
81
Need help?
Do you have a question about the Freecurve and is the answer not in the manual?
Questions and answers