9
|

The lift does not move in one direction

The LED remains dim upon joystick operation in the faulty direction

Connect the short-circuit connectors K6 and K9
A. The lift moves after operation of the direction switch (DS)
The fault is located in the chair
Wire break between K6-3 and the joystick
Wire break between K6-5 and the joystick
Joystick is defect
Earth short at K6-3, K6-5
B. The lift does not move after operation of the direction switch (DS)
The fault is located in the unit
Wire break between K6-3 and K4-11
Wire break between K6-5 and K4-10
Directions switch DS is faulty
Earth short at K4-10, K4-11
10
|
The lift does not move in one direction

The LED flashes upon joystick operation in one direction

Connect the short-circuit connectors K6 and K9
A. The lift moves after operation of the direction switch (DS)
The fault is located in the chair
One of the footrest edges is activated
One of the footrest switches is faulty
The seat frame bottom detection is activated
The seat frame bottom switch is faulty
Wire break between K9-1 and K9-2
Wire break between K9-3 and K9-4
B. The lift does not move after operation of the direction switch (DS)
The fault is located in the unit
One of the cover safety edges is activated
The cover detection switch is faulty
Wire break between K4-4 and K9-3
Wire break between K4-5 and K9-4
Wire break between K4-6 and K9-1
Wire break between K4-7 and K9-2