Handicare Freecurve
5
|
The lift does not move in both directions
The LED lights up upon joystick operation and becomes weak after 3 seconds
Connect the short-circuit connectors K6 and K9
The lift does not move after operation of the direction switch (DS)
•
One of the motors is faulty
(< 2 Ω between K1-3 and K1-4 or K2-7 and K2- 8)
•
One of the brakes is faulty
(resistance shall be 40-70 Ω between K1-1 and K1-2 or K2-5 and K2-6)
6
|
The lift does not move in both directions
The lift is on the charge station. The LED remains dim upon joystick operation in one
direction, in the other direction the Led lights up and gets dimmed, the lift gives one beep.
Connect the short-circuit connectors K6 and K9
A. The lift moves after operation of the direction switch (DS)
The fault is located in the chair
•
Wire break between K6-3 and the joystick
•
Wire break between K6-5 and the joystick
•
Joystick is defect
B. The lift does not move after operation of the direction switch (DS)
The fault is located in the unit
•
Wire break between K6-3 and K4-11
•
Wire break between K6-5 and K4-10
•
Directions switch DS is faulty