1. Safety Precautions
1.3 Safety Precautions
12. ALWAYS use proven disinfection procedures after centrifuging biohazardous materials.
13. Should not centrifuge flammable, toxic, radioactive or explosive materials.
14. When it is necessary to use toxic or radioactive materials or pathogenic micro-organisms
which belong to the Risk Group II of WHO: "Laboratory Bio-Safety Manual," should follow
national regulations.
- Do not place dangerous materials within 30 cm distance around the device.
- Use the emergency lid release function only when the lid button on the
control panel is dumb under the condition of complete stop of rotor running.
- Never try to open or move the device if it is not completely stopped.
- If the power input is more than +/-10% of the recommended voltage or
fluctuates frequently, it may cause malfunction of the device and often
result in serious damage.
- Install the device at the place without any kinds of corrosive gases.
6 / 29