Operating Manual FLETA 5 Clinical Centrifuge Date of Purchase Serial No. Place of purchase...
Page 2
Fleta5 Research Use Only Copyright ⓒ 2018 Hanil Scientific Inc. All rights reserved. Contact Us If you have and questions, contact Hanil Scientific Inc. or place of purchase. +82-2-3472-0727 Inquiry: info@ihanil.com Order: sales@ihanil.com Tech. support: techsupport@ihanil.com The contents in this operating manual are subject to change for device improvement.
CONTENTS General Considerations------------------------------------------------- Safety Transport & Storage Safety label Electric safety information Product description------------------------------------------------------- Structure -5 Delivery package -6 2.3 Technical specifications -7 Installation ---------------------------------------------------------------- Packing Inspection Installation -8 3.2.1 Selecting the location -8 3.2.2 Power connetion -9 Operation ---------------------------------------------------------------- Operation panel 4.2 Opening and closing the lid -11 4.3 Loading and removal of the rotor -11...
2. Check the voltage to be used, before connecting the centrifuge to the power source. 3. Only use rotors, parts, and accessories provided by Hanil Scientific Inc. Hanil Scientific Inc. is not responsible for damages of the device and accidents caused by using parts and accessories not recommended.
1. General Considerations 1.2 Transport & Storage • Only lift and transport the • The device and the accessories device with sufficient number of may only be stored in dry rooms helpers. -Storage- -Transport- Ambient temperature 5℃~35℃ Ambient temperature -10℃~40℃ Maximum relative humidity 30%~85% Maximum relative humidity 10%~90% Air pressure 500~1060hpa...
2. Product description 2.1 Structure ① ② ③ ④ ⑤ ⑥ ⑦ ⑧ ⑨ ①Door: Door protects the inside of the chamber. ②Door latch: Fix the door. ③RPM measuring window: Visual inspection for rotor stop or for a speed check using a stroboscope. ④Chamber: Where the rotor is loaded.
3. Installation 3.1 Packing Inspection •Check packing conditions carefully, before unpacking. •Contact Hani lScientific Inc. immediately if damages found. •Check the delivery for completeness. •You can get contact details on packing boxes and the back of the manual. 3.2 Installation 3.2.1 Selecting the location Installation on hard and flat ground.
3. Installation 3.1 Packing Inspection 3.2.2 Power Connection • Connect the device to voltage sources which correspond to the electrical requirements on the label attached to the device. • Use sockets with a protective earth conductor and provided powercord. 1. Connect the power cable to the power socket on the back side of the product and plug the cord into the outlet.
4. Operation 4.1 Operation Panel ① ② ④ ③ ⑤ ⑥ ⑧ ⑨ ⑦ ⑩ ⑪ ⑫ ⑬ Button Function Turn the dial in the setting mode ① ‘MODE’ dial to move cursor to set the RPM / RCF / Time / Prog./Acc/Dec. ②...
•Closing the lid Put the lid on and press the edge of the lid. 4.3 Loading and removal of the rotor •Only use rotors, parts, and accessories provided by Hanil Scientific Inc. •Do not use scatched or cracked rotor in use. Loading ①...
4. Operation 4.4 Setting rotor/bucket ID • Be sure to set rotor id. or bucket id. correctly before use 1. Press the Enter knob. 2. Turn the Mode knob to move the cursor to rotor ID mode. 3. Turn the Enter knob until the desired rotor/bucket ID displayed. 4.
4. Operation 4.5 Loading tubes • Only use tubes provided or approved by Hanil Scientific Inc. • Always use the same type of tube. • Tubes should be loaded symmetrically. • Do not exceed the maximum rated speed of the tube.
4. Operation 4.6 Setting RPM 1. Press the Enter knob. 2. Turn the Mode knob to select RPM setting mode. ▶ RPM setting mode: RPM flashes. 3. Turn the Enter knob until the desired rpm is displayed. 4. Press the Enter knob to save the value. 5.
4. Operation 4.8 Setting Run Time 1. Press the Enter knob. 2. Turn the Mode knob to select Time setting mode. ▶ Time setting mode: 00:00:00 flashes. 3. Turn the Enter knob until the desired run time is displayed. 4. Press the Enter knob to save the value. 5.
4. Operation 4.10 Setting Deceleration Rate 1. Press the Enter knob. 2. Turn the Mode knob to select DEC setting mode. ▶ DEC setting mode: DEC flashes. 3. Turn the Enter knob until the desired deceleration rate is displayed. 4. Press the Enter knob to save the value. 5.
4. Operation 4.12 Calling Up a Program 1. Turn the Enter knob until desired program number is displayed. ▶ You can see that the program number is changed on the PROG display. 4.13 Starting or Stopping a Run Starting a run 1.
5. Maintenance 5.1 Care instructions • The following procedures should be performed regularly. ①Regularly inspect the rotor chamber for check the motor shaft is nomal. ②Rotate the shaft with your hand to make sure it turns smoothly ③Use the stopwatch to check that the time setting is correct 5.2 Cleaning •...
6. Troubleshooting 6.1 General errors Problem Recommended Action Check the power cord connection. Power failure Check the power fuse of the device. Device cannot be started Check the lid is closed completely. Lid cannot be opened Press the ‘Lid open button’. Remove the dirt at the door latch and Lid cannot be closed close the lid.
- Shut off the power supply, and then, turn on the power switch again to check the device. - Lid open during operation - Shut off the power supply, and contact Hanil open(Operat Scientific Inc. System error - System error - Contact Hanil Scientific Inc.
Page 22
- Shut off the power supply, and then, turn on error the power switch again to check the device. Low voltage - Low Voltage - Contact Hanil Scientific Inc. error RPM Senser - RPM Sensor error - Check if the id disc is on the bottom of the...
6. Troubleshooting 6.3 Emergency lid release In the event of a power failure or the centrifuge lid cannot be opened, active the emergency release. ① Turn off the power switch. ② Wait for the rotor to stop completely. ③ Find the “emergency release key” at the bottom the decvice. ④...