2. Check the voltage to be used, before connecting the device to the power source. 3. Only use rotors, parts, and accessories provided by Hanil Scientific Inc. Hanil Scientific Inc. is not responsible for damages of the device and accidents caused by using parts and accessories not recommended.
1. General Considerations 1.2 Transport & Storage -Storage- -Transport- Ambient temperature -10℃~40℃ Ambient temperature 5℃~40℃ Maximum relative humidity 10%~90% Maximum relative humidity 30%~85% Air pressure 500~1060hpa Air pressure 500~1060hpa 1.3 Safety label on the Device Caution/Warning Warning for electric shock Insert tube symmetrically.
2. Product Description 2.1 Structure ① ② ③ ④ ⑤ ⑥ ⑦ ⑧ ① Door ② RPM measuring window ③ Chamber ④ Ventilation slots ⑤ Display & control panel ⑥ Power Socket ⑦ Drain cap ⑧ Power Swithch 2.2 Delivery package - Device - Power cable - Operating manual...
Dimension (W x D x H, mm) 285.2 x 593.3 x 281 Weight without rotor 33 kg Power requirement (VA) 2.5 kVA Power input (V, Hz) 220V 50/60Hz Cat. No. M15R Technical Support : +82-3452-8966 / techsupport@ihanil.com 7 / 22...
3. Installation 3.1 Packing Inspection • Check packing conditions carefully, before unpacking. • Contact Hanil Scientific Inc. immediately if damages found. • Check the delivery for completeness. ▶ You can get contact details on the packing box or manual. 3.2 Installation Installation on hard and flat ground.
3. Installation 3.3 Power Connection • Connect the device to voltage sources which correspond to the electrical requirements on the label attached to the device. • Use sockets with a protective earth conductor and provided power cord. 1. Connect the power cable to the power socket on the right side of the device and plug the cord into the outlet.
• Since this product is equipped with a rotor, no additional installation work is required. • Use only rotors produced by Hanil Scientific Inc. • Do not use rotors that are corroded or have mechanical defects. • Be sure to install the rotor lead before use.
Page 11
3. Installation 3.6 Rotor ID. • For this machine the user must input the rotor ID manually. • The rotor ID must be entered correctly so that it can be centrifuged within the maximum speed. 1. Press the Setup/Enter button. 2.
3. Installation 3.7 Loading Tubes • Only use tubes provided or approved by Hanil Scientific Inc. • Always use the same type of tube. • Tubes should be loaded symmetrically. • Do not exceed the maximum rated speed of the tube.
4. Operation 4.1 Control Panel ① ⑥ ⑦ ⑤ ⑧ ③ ④ ⑨ ② Button Description ① Up the values when setting parameters Down Down the values when setting parameters ② /Fast cool /Pre-cooling the rotor and chamber ③ Left Left button is used to select parameter ④...
4. Operation 4.2 Setting RPM/RCF 1. Press the Setup/Enter button. 2. Press the Left or Right button until the RPM/RCF display(rpm) flashes. ▶ RPM setting mode: RPM(rpm) display flashes. ▶ RCF setting mode: RCF(xg) display flashes. 3. Press Up or Down button until the desired RPM/RCF is displayed. 4.
4. Operation 4.4 Setting Temperature 1. Press the Setup/Enter button. 2. Press the Left or Right button until the 'Temperature' display(-00℃) flashes. 3. Press Up or Down button until the desired temperature is displayed. 4.5 Fast Cool If the samples are "temperature sensitive", it is critical to pre-cool the centrifuge and the rotor.
4. Operation 4.8 Short Run After reaching the set speed, the centrifugation is decelated and stopped 1. Set the desired speed. ▶ Refer to [4.2 Setting Speed] 2. Press the Short button more than 2sec. 4.9 Saving/Calling up a Program [Saving] 1.
5. Maintenance 5.1 Care Instructions 1. Regularly inspect the rotor chamber for check the motor shaft is normal. 2. Regularly check the rotor and accessories to be sure there is no damage. 3. Rotate the motor shaft with your hand to make sure it turns smoothly. 4.
6. Troubleshooting 6.1 Possible Problems Problems Recommended Action - Check the power cord connection. Power failure - Check the power fuse of the device. Centrifugation - Check the lid is closed completely. cannot be started Door cannot be - Press the ‘Door' button. opened - Check the power connection.
6. Troubleshooting 6.2 Error Codes Code Message Cause Recommended Action • Check the tightness of rotor loading. • Asymmetrical arrangement of Imbalance • Check the proper loading of tubes. tubes. error • Check the stable ground or • Different volume of sample . worktable.
6. Troubleshooting 6.3 Emergency Door Open If the device door cannot be opened, you can activte the emergency open manually. 1. Wait for rotor to stop before activating the emergency open. 2. Insert the T-wrench into the emergency open hole and turn it clockwise until the door is opened.
Page 22
7. Rotor 7.1 Rotor Rotor Tube Required Bore Ø x L (mm) Max. RPM(rpm) Capacity Adaptor Radius (mm) Max. RCF (xg) Hole angle : ∠45° 0.2mL 6.5 x 17 12,500 Max. Capacity : 32 x 0.2 mL 0.2mL strips 55.5 Size (ø...
Page 24
RECYCLABLE 16 Arayukro, Gimpo 10136, Rep.of Korea Service Inquiries: +82-2-3452-8966 techsupport@ihanil.com www.ihanil.com Doc.No. : OMM9PDEN2002...