Page 2
LOW SPEED CENTRIFUGE T04B - Product name :Centrifuge - Model name : T04B - Manufacturer : Hanil Scientific Inc. - Address : 16 Arayukro, Gimpo, #10136, Korea Research Use Only www.ihanil.com UM-T04B(E)(Rev.1), 20210406...
USER MANUAL T04B 1. Safety warnings and cautions 1.1. Safety label This manual describes the performance, usage, and handling precautions of the purchased product. Please read carefully before using the product. 1.2. Conditions of use, storage or transport Rotor insertion, tightning caution...
3. Use only the rotor and recommended parts and accessories provided by Hanil Scientific Inc. We are not responsible for any damage to the device or accidents resulting from the use of non- recommended parts and accessories.
USER MANUAL T04B 2. Product composition and information 2.1. Appearance Power socket & fuse Power switch Display & Control Manual Lid Open Hole Rotor - optional 2.2 Components Components Grease User manual AC Power Cord Rotor Locking Emergency Lid Tool...
3. Product installation 3-1. Unpacking 1. After purchasing the centrifuge, open the box and check the components. Centrifuge (T04B) / User Manual / AC power cord / Emergency lid open tool / Rotor locking tool / Grease 3-2. Power connection 1.
USER MANUAL T04B 3.3 Lid open Used to open the lid. When Lid is closed, the lamp is off, and when it is open, the lid lamp is turned on. 1. When the lid is closed (Lid lamp is off), press [Open].
Before use, make sure that the rotor is firmly connected to the motor shaft. 3.5 Tube mounting · Use the tube for centrifuge recommended by Hanil Scientific Inc, and do not use it above the allowable standard after checking the maximum RCF value for each tube.
USER MANUAL T04B 2. Sample tubes must be placed and mounted symmetrically. ▶ The tube filled with the sample, there should be no difference in weight and the density should be symmetrically and evenly arranged. ▶ Tubes for centrifuges must be used. Max. It should be used after checking the RCF value.
USER MANUAL T04B 4. How to use 4.1 Control Panel 4000 99:59 9 9 □ RPM/RCF The rotation speed is indicated in RPM/RCF, and mutual conversion is possible by pressing the RPM / RCF button. Control interval : 1 □ Pulse The pulse button is for when press &...
USER MANUAL T04B 4.2 Speed setting The speed setting value is displayed as RPM and RCF, and is automatically calculated through interlocking. 1. Press [RPM/RCF] ▶ Press 1/2 → RPM/RCF setting mode 4000 99:59 9 9 ▶ Entering the setting mode, RPM or RCF is displayed on the display screen.
Page 15
USER MANUAL T04B Time mode definition (ALL mode / At set Speed mode) For accurate time management, the time mode can be set to ALL mode (time increases with start) and At set Speed mode (time increases after reaching the set speed).
USER MANUAL T04B 4.4 Acceleration/deceleration (ACC/DEC) setting Acceleration speed can be set up to 9 steps and deceleration speed up to 10 steps (natural deceleration:0) for sensitive sample protection and clean layer separation. 1. Press [ACC] to set acceleration of [DEC] to set deceleration.
USER MANUAL T04B 4.6 Start/stop Can be used to start or stop motion. During operation, the Start/Stop lamp is indicated by lighting. 4.6.1 Start 1. After setting the speed and time, etc., press the [Start/Stop] button. 4000 99:59 9 9 ▶The [Start/Stop] lamp lights up during operation.
USER MANUAL T04B 4.7 Change setting value during operation 1. If you want to change the setting value during operation, press the mode button and enter the setting value to be changed. ▶Speed / time / ACC / DEC setting value can be changed during operation.
USER MANUAL T04B Program call: It is designed to call the program stored between 00 and 99. 1. Press [PROG] once. ▶Display→ CALL flashing 4000 99:00 9 9 2. Enter the program number to be called and press [Enter]. ▶Press [Enter] → Program call 4000 99:00 9 9 ▶The set values (speed, time, acc/dec, etc.)
USER MANUAL T04B 4.11 Set the number of end tones You can adjust the number of ending sounds that sound at the end of the operation up to 99 (0~99, silent: 0). 1. Press and hold [AT SET SPEED] for 2 seconds or longer.
USER MANUAL T04B 5. Maintenance Main body 1. If the exterior is contaminated, wash it with soapy water on a soft cloth and wipe it with a dry cloth to prevent moisture from remaining. 2. Do not use chemicals such as alcohol, benzene, benzol, or thinner as they may damage it.
USER MANUAL T04B 6. Problem solving 6.1 Checklist before reporting a malfunction If there is a problem with the centrifuge, check the following before requesting a service center. Symptom Check list [Refer to [3.2 Power Connection] and check if the power plug No power is removed.
USER MANUAL T04B 6.2 Error message information If the problem persists after taking the following measures, please contact the service center. Error code Description Action Motor start error: Occurs when the motor Error 1 Check the rotor rotation through the lid's center window.
USER MANUAL T04B Rotor & Accessories Angle Rotor, - Capacity : 10 x 50 mL or 50 mL Conical - Max. RPM / RCF : 4,000 / 2,549 - Hole angle rotation : ∠ 45° AO-50-10 - Hole dimension (Ø x L,mm) : 32.2 x 13.5...
Page 25
USER MANUAL T04B Tube 15 mL conical 25mL conical 25mL conical 50 mL conical Tube capacity (mL) 17 x 120 28.8x83 28.8x78.5 25.7 x 101.4 29 x 108 29.5 x 118 Tube Dimension(Φ x L,mm) None Adapter TR15c(50c) TR25c(50c) TR25c(50c)
Page 26
USER MANUAL T04B Swing Rotor, - 4 loadings SO-100-4 - Max. RPM : 4,000 - Angle from axis during rotation: ∠ 90° - Rotor dimension / weight (Ø x L,mm/g) : 206.3 x 46 / 810 - Supplied with a lubricant - No ID ring - Max.
Page 27
USER MANUAL T04B Max. height for tube fit (mm) : 125 50 mL Conical Bucket, Bd50c Hole bottom type : Conical Max. RPM / RCF 4 : 4,000 / 2,774 Max. Radius (mm) : 155.1 Hole dimension (Ø x L, mm) : 30 x 89...
Page 28
USER MANUAL T04B 15 mL Bucket, B15-8 - Hole bottom type : Flat - Max. RPM / RCF : 4,000 / 2,719 - Max. height for tube fit (mm) : 115 - Max. Radius (mm) : 152 - Supplied with 3.0 mm thick NBR rubber pad - Hole dimension (Ø...
Page 29
USER MANUAL T04B - Max. height for tube fit (mm) : 125 150 mL Conical Bucket, B50c - Hole bottom type : Conical - Max. RPM / RCF : 4,000 / 2,700 - Max. Radius (mm) : 150.9 - Hole dimension (Ø x L,mm) : 29.5 x 91.5...
Page 30
USER MANUAL T04B Swing Rotor, - 6 loadings SO-15-6 -Max. RPM : 4,000 -Angle from axis during rotation: ∠ 90° -Rotor dimension / weight (Ø x L,mm / g) : 198.2 x 44 / 573 -No ID ring 15 mL Dual Round Bucket, Bd15 - Max.