Wrong voltage can damage equipment. 3. Use rotors provided by Hanil Science Industrial Co Ltd and use parts and accessories recommended. Hanil Science Industrial Co Ltd is not responsible for damages of equipment and accidents caused by using parts and accessories not recommended.
2.1 Packing Inspection · Check packing conditions very carefully after arrival of centrifuge. · Contact Hanil Science Industrial Co Ltd immediately if damages found . · You can get contact details on packing boxes and outer cover of manual. 2.2 Installation directions ·...
2. Installation Directions 2.3 Rotor · Rotor, tube rack(adapter), tube and adapter set recommended by manufacturer should be used. · Do not use rotor and bucket with a visible corrosion or mechanical fault. · Waring! In danger of injury due to asymmetric installation of a rotor. ·...
Page 7
2. Installation Directions Tubes arrangement Same volume of sample should be put on tubes after you measure it exactly. Tubes should be placed on a rotor symmetrically. At this moment, if volume of symmetrical tube is different, a rotor and an axis of rotation can be damaged due to severe vibration during rotation.
⑨ Emergency Lid Open system: This is where you can open a lid manually in case of emergency such as power failure ⑩ Drain : This is where sample or water flowed into the chamber is discharged 3.2 Components ① Smart R17 Main body ② Power cable ③ User manual ④ Rotor Locking Tool (T wrench) ⑤...
0.2ml x 32 (PCR Rotor) Time 99min 59sec ACC ramps 6 steps DEC ramps 6 steps Power supply AC 220V, 60Hz Dimension (W x D x H, mm) 281 x 595 x 285 mm Weight (kg) 30 (Without rotor) Cat. No. Smart R17 www.ihanil.com...
3. Inspection of operation 3.4 Controller ② ① ⑦ ⑥ ④ ③ ⑧ ⑨ ⑤ Button Function ① Display Set-up information, operating condition is displayed. This is to select program and to increase set value of program as the upper button of ②...
3. Inspection of operation 3.5 RPM/RCF Input 1.In standby mode, press Setup/Enter ( ) button. 2. Press Left ( ) or Right ( ) button to change to rpm or rcf mode 3. Press Up ( ) or Down ( ) button to input the setting value.
3. Inspection of operation 3.8 Temperature Setting 1.In standby mode, press Setup/Enter ( ) button. 2. Press Left ( ) or Right ( ) button to move to TEMP position. 3. Press Up ( ) or Down ( ) button to input the setting value. ▶Setting temperature is available from -10℃...
3. Inspection of operation 3.11 Saving and Recalling Program Saving programs 1. Press Call/Save ( ) button. 2. Select program No. by pressing Up ( ), down ( ) button firstly and then, press Call/Save ( ) button. 3. Press the Setup/Enter ( ) button to set menus.
4. Cleaning and Maintenance 4.1 General check points Please check it out for followings 1. Shaft : Do not bend a shaft. Check whether they are worn out. 2. Motor bearing : check whether it makes noise or rotates smoothly. 3.
Page 15
4. Cleaning and Maintenance Rotor 1. Corrosion of rotor and surface treatment If a rotor is exposed to acid or basic liquid, it gets damaged. In case rotor weight is imbalanced due to corrosion, serious vibration and noise can happen when high speed rotating. Furthermore, it can damage a drive shaft.
• Over the pre set limit. (The value of L-Cal of the preset menu) Temp. • Temp sensor connector has not • Check the connection of mainboard Sensor connected. connector. error * Not satisfied with the above troubleshooting, contact Hanil customer service center. www.ihanil.com...
5. Troubleshooting 5.2 Emergency lid open system In case a lid closed due to power failure, activate emergency lid open system manually. 1.Switch off main power. Wait till a rotor stops through RPM measuring window. 2.As the below picture, Insert a flathead screwdriver into emergency lib open system on the left side of the equipment.