Page 1
Operation Manual FLETA 4 Clinical Centrifuge Date of Purchase Serial No. Place of purchase...
Page 2
Fleta4 Research Use Only Copyright ⓒ 2018 Hanil Scientific Inc. All rights reserved. Contact Us If you have and questions, contact Hanil Scientific Inc. or place of purchase. +82-2-3472-0727 Inquiry: info@ihanil.com Order: sales@ihanil.com Tech. support: techsupport@ihanil.com The contents in this operating manual are subject to change for device improvement.
2. Check the voltage to be used, before connecting the centrifuge to the power source. 3. Only use rotors, parts, and accessories provided by Hanil Scientific Inc. Hanil Scientific Inc. is not responsible for damages of the device and accidents caused by using parts and accessories not recommended.
1. General Considerations 1.2 Transport & Storage • Only lift and transport the • The device and the accessories device with sufficient number of may only be stored in dry rooms helpers. -Storage- -Transport- Ambient temperature 5℃~ 40℃ Ambient temperature -10℃~40℃ Maximum relative humidity 30%~85% Maximum relative humidity 10%~90% Air pressure 500~1060hpa...
2. Product description 2.1 Structure ① Lid ② Lid Latch ③ RPM measuring window: Visual inspection for rotor stop or for a speed check using a tachometer. ④ Chamber: Where the rotor is loaded. ⑤ Lid Packing ⑥ Rotor ⑦ Display&Control panel: The display shows time, rpm, temperature and etc. ⑧...
3. Installation 3.1 Packing Inspection •Check packing conditions carefully, before unpacking. •Contact Hanil Scientific Inc. immediately if damages found. •Check the delivery for completeness. •You can get contact details on packing boxes and the back of the manual. 3.2 Installation 3.2.1 Selecting the location...
3. Installation 3.3 Power Connection •Connect the device to voltage sources, which correspond to the electrical requirements on the label attached to the device. •Check the wall outlet is earth-grounded, surge-protected and can supply enough power. 1. Connect the DC plug at the power socket on the right side of the device. 2.
Operation 4.1 Operation Panel Button Description Setup / Enter Set the Values Up or down the values UP / Down Move(Left/Right) Select a parameter Start/Stop the centrifugation Start/Stop/Short Press more than 2sec. : short spin Lid open Press once: Saving a program (Up to 10) Program Press twice: Calling up a program www.ihanil.com...
Operation 4.2 Opening/Closing the lid •The lid can only be opened if the centrifuge is switched on. •Do not reach with your fingers between the housing and lid. •Close the lid completely before operation. Opening the Lid 1. Press the Lid button( Closing the Lid 1.
Operation 4.3 Loading/Unloading a rotor •Only use rotors, parts, and accessories provided by Hanil Scientific Inc. •Do not use scatched or cracked rotor in use. Loading ① Clean the motor shaft and the chamber. - When loading the rotor, the temperature should be 10 to 30 degrees.
4. Operation 4.4 Loading tubes • Only use tubes provided or approved by Hanil Scientific Inc. • Always use the same type of tube. • Tubes should be loaded symmetrically. • Do not exceed the maximum rated speed of the tube.
4. Operation 4.5 Setting Rotor Id. 1.Press the Setup/Enter ( ) button. 2. Press the Left( ) or Right( ) button to select rotor Id. mode. ▶Rotor Id. setting mode : Rotor flashes on display 3. Press the up( ) or down ( ) buttons until the desired rotor id displayed.
4. Operation 4.10 Saving/Calling up a program Saving a program ① Set the centrifugation time/speed/Accel/Decel values. ② Press Program ( ) button once. ▶You can store 10 programs on the device: 0~9 ③ Select the program number by pressing the Up( ) or Down( ) button.
②Rotate the shaft with your hand to make sure it turns smoothly ③Use the stopwatch to check that the time setting is correct ④If you find any damages, do not use the device. Contact Hanil Scientific Inc. 5.2 Cleaning • Before cleaning the centrifuge, be sure to switch off the device and disconnect the power cord.
6. Troubleshooting 6.1 General errors Problem Recommended Action Check the power cord connection. Power failure Check the power fuse of the device. Device cannot be started Check the lid is closed completely. Lid cannot be opened Press the ‘Lid open button’. Remove the dirt at the door latch and close the lid.
- Shut off the power supply, and then, turn on the power switch again to check the device. - Lid open during operation - Shut off the power supply, and contact Hanil open(Operat Scientific Inc. System error - System error - Contact Hanil Scientific Inc.
Page 21
- Shut off the power supply, and then, turn on error the power switch again to check the device. Low voltage - Low Voltage - Contact Hanil Scientific Inc. error RPM Sensor - RPM Sensor error - Check the id disc is on the bottom of the...
Page 22
6. Troubleshooting 6.2 Error codes If the centrifug lid cannot be opened, you can activte the emergency open manually. 1. Turn off the power switch and wait for rotor to stop before activating the emergency open. Insert the provided T-wrench into the emergency release hole and turn it clockwise until the lid is released.