3. Installation

3.7 Positioning of Sample Tubes

- If it is hard to make balance with the samples, please employ control
tubes. Otherwise, it may bring about noise or vibration, which may lead
to damaging the device.
Correct Arrangement
Wrong Usage of Tubes
Different Caps
1. Before loading sample tubes, check the water drop or dirt in the rotor hole or inner sleeve.
▶ If you find water drops or dirt in the rotor or inner sleeve, remove it with soft dry cloth.
2. Tubes should be placed in the rotor with equal amount of samples at symmetrical
positions.
▶ Only use appropriate centrifugal tubes and do not exceed the speed beyond the max
g-force allowed for the tubes.
14/29
www.ihanil.com
Different Volume
Different Density
Different Tubes