CONTENTS General Considerations------------------------------------------------- Safety Transport & Storage Safety label Electric safety information Product description------------------------------------------------------- Structure -5 Delivery package -6 2.3 Technical specifications -7 Installation ---------------------------------------------------------------- Packing Inspection Installation -8 3.2.1 Selecting the location -8 3.2.2 Power connetion -9 Operation ---------------------------------------------------------------- Operation panel 4.2 Opening and closing the lid -11 4.3 Loading and removal of the rotor -11...
2. Check the voltage to be used, before connecting the centrifuge to the power source. 3. Only use rotors, parts, and accessories provided by Hanil Science Inc. Hanil Scientific Inc. is not responsible for damages of the device and accidents caused by using parts and accessories not recommended.
1. General Considerations 1.2 Transport & Storage • Only lift and transport the • The device and the accessories device with sufficient number of may only be stored in dry rooms helpers. -Storage- -Transport- Ambient temperature 5℃~35℃ Ambient temperature -10℃~40℃ Maximum relative humidity 30%~85% Maximum relative humidity 10%~90% Air pressure 500~1060hpa...
2. Product description 2.1 Structure ① ② ③ ④ ⑤ ⑥ ⑦ ⑧ ⑨ ①Door: Door protects the inside of the chamber. ②Door latch: Fix the door. ③RPM measuring window: Visual inspection for rotor stop or for a speed check using a stroboscope. ④Chamber: Where the rotor is loaded.
3. Installation 3.1 Packing Inspection •Check packing conditions carefully, before unpacking. •Contact Hani lScientific Inc. immediately if damages found. •Check the delivery for completeness. •You can get contact details on packing boxes and the back of the manual. 3.2 Installation 3.2.1 Selecting the location Installation on hard and flat ground.
3. Installation 3.1 Packing Inspection 3.2.2 Power Connection • Connect the device to voltage sources which correspond to the electrical requirements on the label attached to the device. • Use sockets with a protective earth conductor and provided powercord. 1. Connect the power cable to the power socket on the right side of the product and plug the cord into the outlet.
4. Operation 4.1 Operation Panel ① ② ④ ③ ⑤ ⑥ ⑧ ⑨ ⑦ ⑩ ⑪ ⑫ ⑬ Button Function Turn the dial in the setting mode ① ‘MODE’ dial to move cursor to set the RPM / RCF / Time / Prog./Acc/Dec. ②...
•Closing the lid Put the lid on and press the edge of the lid. 4.3 Loading and removal of the rotor •Only use rotors, parts, and accessories provided by Hanil Scientific Inc. •Do not use scatched or cracked rotor in use. Loading ①...
4. Operation 4.4 Loading tubes • Only use tubes provided or approved by Hanil Scientific Inc. • Always use the same type of tube. • Tubes should be loaded symmetrically. • Do not exceed the maximum rated speed of the tube.
4. Operation 4.5 Setting RPM/RCF ① Press the Mode ( ) button in stand-by mode. ②Press the Enter ( ) button to set RPM. ▶ Press the Enter button once more to set RCF. - RPM mode: RPM flickering - RCFmode: RCF flickering ③...
4. Operation 4.7 Acc/Dec setting ① Press the Enter ( ) button in stand-by mode. ② Turn the Mode ( ) button to move the cursor to ‘ACC’ or ‘DEC’. ③ Turn the Enter ( ) button to the left or right to input the value. ④...
4. Operation 4.11 Saving / Recall a program Saving a program ① Set the centrifugation time/speed/Accel/Decel values. ② Press Enter ( ) button. ③ Turn Enter ( ) button to set Program No. ④ Turn Move ( ) button to select [YES] ⑤...
5. Maintenance 5.1 Care instructions • The following procedures should be performed regularly. ①Regularly inspect the rotor chamber for check the motor shaft is nomal. ②Rotate the shaft with your hand to make sure it turns smoothly ③Use the stopwatch to check that the time setting is correct 5.2 Cleaning •...
6. Troubleshooting 6.1 General errors Problem Recommended Action Check the power cord connection. Power failure Check the power fuse of the device. Device cannot be started Check the lid is closed completely. Lid cannot be opened Press the ‘Lid open button’. Remove the dirt at the door latch and Lid cannot be closed close the lid.
Page 19
6. Troubleshooting 6.2 Error code Error Display Cause Solution Check the power cable - No power connection. No display Check the main fuse - Failure of power supply None in laboratory and device - No rotor - Load a rotor. - Turn off the main power and Errors in drive or rotor recognition function.
6. Troubleshooting 6.3 Emergency lid release In the event of a power failure or the centrifuge lid cannot be opened, active the emergency release. ① Turn off the power switch. ② Wait for the rotor to stop completely. ③ Find the “emergency release key” at the bottom the decvice. ④...