LOW SPEED CENTRIFUGE - Product name :Centrifuge - Model name : T04 - Manufacturer : Hanil Scientific Inc. - Address : 16 Arayukro, Gimpo, #10136, Korea Research Use Only www.ihanil.com UM-T04(E)(Rev.0), 20210121...
USER MANUAL 1. Safety warnings and cautions 1.1. Safety label This manual describes the performance, usage, and handling precautions of the purchased product. Please read carefully before using the product. 1.2. Conditions of use Rotor insertion, tightning caution Lid closure caution indication warning indication warning In case of Emergency, the Bucket mounting...
3. Use only the rotor and recommended parts and accessories provided by Hanil Scientific Inc. We are not responsible for any damage to the device or accidents resulting from the use of non- recommended parts and accessories.
USER MANUAL 2. Product composition and information 2.1. Appearance Power Socket & Fuse Power Swich Manual Lid Opening Hole Display & Control T04 is provided with AO-15/10-12 Rotor 2.2 Components Components Grease User manual AC Power Cord Rotor Locking Emergency Lid...
3-1. Unpacking 1. After purchasing the centrifuge, open the box and check the components. Centrifuge (T04) / Rotor ( AO-15 / 10-12 ) / User Manual / AC power cord / Emergency lid opening tool / Rotor locking tool / Grease 3-2.
USER MANUAL 3.3 Lid open Used to open the lid. When the Lid is closed, the lamp is off, and when it is open, the lamp is turned on. 1. When the lid is closed (The lamp is off), press [Open]. ▶...
Before use, make sure that the rotor is firmly connected to the motor shaft. 3.5 Tube mounting · Use the tube for centrifuge recommended by Hanil Scientific Inc, and do not use it above the allowable standard after checking the maximum RCF value for each tube.
Page 12
USER MANUAL 2. Sample tubes must be placed and mounted symmetrically. ▶ The tube filled with the sample should have no difference in weight and the density should be symmetrically and evenly arranged. ▶ Tubes for centrifuges must not exceed Max RCF value. It should be used after checking the RCF valueof tubes.
USER MANUAL 4. How to use 4.1 Control Panel □ RPM/RCF The rotation speed is indicated in RPM/RCF, and mutual conversion is possible by pressing the RPM / RCF button. Control interval : 1 □ Time The time can be set in “minute” and “second”, and up to 99 minutes 59 seconds or continuous operation is possible.
USER MANUAL 4.2 Speed setting The speed setting value is displayed as RPM and RCF, and is automatically calculated through interlocking. 1. Press [RPM/RCF] ▶ Press Once/Twice → RPM/RCF setting mode ▶ Entering the setting mode, RPM / RCF lamp will turn on. 2.
USER MANUAL 4.3 Time setting The time can be set in "minute" and "second", up to 99 minutes 59 seconds or continuous operation (00 minutes 00 seconds setting) is possible. 1. Press [Time]. ▶pressing [Time] Once/Twice → "Minuite"/"Second" input mode ▶Entering the setting mode, Min / Sec lamp will turn on.
USER MANUAL 4.4 Start / Stop Can be used to start or stop operation. During operation, the Start/Stop lamp is indicated by lighting up. 1. Start 1. After setting the speed and time, etc., press the [Start/Stop] button. ▶Operation starts only when the lid is closed. ▶During operation →...
USER MANUAL 4.5 program storage Program storage: When operating the device under various conditions, setting values such as speed and time can be saved in advance and then recalled and used immediately as needed. 1. After setting the program that needs to be saved, press the [PROG] button for more than 3 seconds.
USER MANUAL 4.6 Program call Program call: It is designed to call the program stored between 00 and 10. 1. Press [PROG] once. ▶Display→ CALL flashing 2. Using the [▲/▼] key, select the program number and press [PROG] again. ▶Selected program number will show the saved settings ▶If you do not press [PROG] for 5 seconds, the setting mode is...
USER MANUAL 5. Maintenance Main body 1. If the exterior is contaminated, wash it with soapy water on a soft cloth and wipe it with a dry cloth to prevent moisture from remaining. 2. Do not use chemicals such as alcohol, benzene, benzol as they may damage it. 3.
USER MANUAL T04 6. Problem solving 6.1 Checklist before reporting a malfunction If there is a problem with the centrifuge, check the following before reaching to a service center. Symptom Check list Refer to [3.2 Power Connection] and check if the power plug No power is removed.
USER MANUAL T04 6.2 Error message information If the problem persists after taking the following measures, please contact the service center. Error code Description Action Motor start error: Occurs when the motor Error 1 Check the rotor rotation through the lid's center window.