Operating Manual
Micro 12
Micro Centrifuge

Advertisement

Table of Contents
loading

Summary of Contents for Hanil Micro 12

  • Page 1 Operating Manual Micro 12 Micro Centrifuge Purchase Date Serial No. Place of purchase...
  • Page 2 Copyright ⓒ 2017 Hanil Scientific Inc. All rights reserved. Contact Us If you have any questions, contact Hanil Scientific Inc. or place of purchase. techsupport@ihanil.com / info@ihanil.com 2/20 www.ihanil.com...
  • Page 3: Table Of Contents

    CONTENTS 1. General Considerations -----------------------4 1.1 Safety -4 1.2 Transport & Storage -4 1.3 Safety labe -5 1.4 Electric safety information -5 2 . Product description --------------------------6 2.1 Appearance -6 2.2 Delivery package -6 2.3 Specifications -7 3 . Installation -------------------------------------8 3.1 Packing inspection -8 3.2 Installation -8 3.3 Connect power -9...
  • Page 4: General Considerations

    2. Check the voltage to be used, before connecting the centrifuge to the power source. 3. Only use rotors, parts, and accessories provided by Hanil Scientific Inc. Hanil Scientific Inc. is not responsible for damages of the device and accidents caused by using parts and accessories not recommended.
  • Page 5: Electric Safety Information -5

    1. General Considerations 1.3 Safety label attached to products Electric shock Do not open lid until rotor stops completely. Rotation direction of the rotor. 1.4 Electric safety information 1. Use a power cord only provided with the device. 2. Use sockets with a protective earth conductor and suitable power cord. 3.
  • Page 6: Product Description

    2. Product description 2.1 Appearance ④ ① ② ③ ④ ⑤ ⑦ ⑥ ⑧ ⑨ ① Lid : Lid protects the inside of the chamber. ② Rotor locking screw: Connect the rotor to the motor shaft. ③ Rotor : Connected to the motor shaft for centrifugal separation ④Lid sensor : Detect whether lid is cloesd ⑤...
  • Page 7 2. Product description 2.3 Specifcations Max. RPM 13,500 rpm Max. RCF 12,032 xg Time ≤ 15 min, Pulse Acc. time to max. speed ≤18 sec Dec. time from max. speed ≤8 sec Tube Capacity 1.5/2.0ml x 12 Power supply AC110/220V, 50/60Hz Dimension (W x D x H, mm) 190 x 202 x 129 Weight (kg)
  • Page 8: Installation

    3. Installation 3.1 Packing inspection · Check packing conditions very carefully before unpacking. · Contact Hanil Scientific immediately if damages found. · Check the delivery for completeness. · You can get contact details on packing box or back of the manual.
  • Page 9: Installation -8

    3. Installation 3.3 Connect power · Connect the device to voltage sources which correspond to the electrical requirements on the label attached to the device. · Use sockets with a protective earth conductor and provided power cord. 1. Connect the power cable to the power socket on the right side of the product and plug the cord into the outlet.
  • Page 10: Loading And Removal Of The Rotor -10

    3.5 Loading and removal of the rotor · This device is equipped with a rotor . · Only use rotors, parts, and accessories provided by Hanil Scientific Inc . · Do not use a defective rotor . Follow the steps below to remove or replace the rotor.
  • Page 11: Loading Tubes -11

    3. Installaion 3.6 Loading tubes · Only use tubes provided or approved by Hanil Scientific Inc . · Always use the same type of tube . · Tubes should be loaded symmetrically . · Do not exceed the maximum rated speed of the tube .
  • Page 12 4. Operation 4.1 Operation panel 버튼 설명 ① RPM setting knob Rotate clockwise or counterclockwise to set the rpm value. Rotate clockwise or counterclockwise to set the run time. ② Time setting knob (To use pulse function, set the time to '0' and press the start key.) ③...
  • Page 13: Setting Rpm -13

    4. Operation Setting Rotate the 'RPM setting knob' clockwise or counterclockwise to set the rpm value. ▶When using a A0.2-32 (PCR strip rotor) do not exceed 6,000 rpm. 4.3 Setting run time Rotate 'Time setting knob' clockwise or counterclockwise to set the run time. ▶0~15 min.
  • Page 14: Maintenance

    5.Maintenance Care instructions • The following procedures should be performed regulary. 1. Regularly inspect the rotor chamber for check the motor shaft is nomal. 2. Rotate the shaft with your hand to make sure it turns smoothly. 3. Use the stopwatch to check that the time setting is correct Cleaning •...
  • Page 15: Troubleshooting

    6.Troubleshooting 6.1 Possible problems Please check the following information, if the device is faulty. 증상 확인 사항 전원이 들어오지 않아요. [3.3 전원연결]을 참조하여 전원플러그가 빠져 있는지 확인해주십시오. Lid가 닫혀지지 않을 경우 동작되지 않습니다. [3.4 Lid 열기]를 동작되지 않아요. 참조하여 lid상태를 확인하고 lid를 잘 닫아주십시오. Lid 개방...
  • Page 16 6. Troubleshooting 6.2 Possible problems Problem Recommended Action 1. The device does not operate. Need to repair the controller. 2. The device stops during opertaion. Please contact your dealer. 3. There is no beep sound when power is connected. 1. It operates when the lid is open. Lid sensor replacement may be required.
  • Page 17: Rotor And Accessories

    7. Rotor and accessories Tube Tube capacity 0.2 ml 0.5 ml 1.5/2.0 ml Adaptor None Fixed Angle Rotor, A2.0-12 Cat. No. TR0.2 TR0.5 Adaptor bore 12 x 1 . 5 / 2 . 0 ml 6 x 21 8 x 37 (Фx L, mm) Radius (mm) Hole diameter ( m m) : 11.5...