10.8 Mechanical requirements
Operation
Vibrations
Shock
Packed unit
Vibrations
Shock
10.9 Safety specifications
Safety standards
Protection class
Degree of protection according to DIN 40050
Conformity
CAUTION
Subsequent installation of a stand
The SCD / SMD 19102 C and SCD / SMD 19102 CP monitors are delivered without a
stand. To ensure that a system comprising a monitor and base meet the requirements of
standard EN 60601, select a stand for subsequent installation that meets this standard.
SCD 19102 / SMD 19102
Instructions for Use, 08/2015
10.8 Mechanical requirements
According to EN 60068-2-6
10 ... 58 Hz at ± 0.075 mm deflection
58 ... 500 Hz at 10 m/s²
According to EN 60068-2-27 (single impact)
150 m/s², 6 ms
Under operating conditions, the device must not
be exposed to continuous or repeated shocks.
According to EN 60068-2-6
5 ... 9 Hz at ± 3.5 mm deflection
9 ... 500 Hz at 10 m/s²
According to EN 60068-2-27 (single impact)
250 m/s², 6 ms (in storage packaging)
to EN 60068-2-29 (continuous shocks)
•
IEC 60601-1 +A1:1991 + A2:1995
•
IEC 60601-1:2005 +CORR.1 (2006) +
CORR. 2 (2007)
•
CAN/CSA C22.2 No. 601.1-M90
•
CAN/CSA C22.2 No. 60601-1:08
•
UL60601-1, 1st edition
•
ANSI/AAMI ES60601-1:2005 / A2:2010
Protection class I
IP20
CE according to MDD 93/42/EEC (Class I)
Technical specifications
75