Eizo SCD 19102 Instructions For Use Manual page 41

1.3 mp 19" lcd monitor
Hide thumbs Also See for SCD 19102:
Table of Contents

Advertisement

Servicelevel 2
Calibration
SCD 19102 / SMD 19102
Instructions for Use, 08/2015
Function
Adjustment /
setting range
Information
Temperature
Backlight com-
Information
mand
Backlight sen-
Information
sor
info
When backlight
0 ... 100 %
regulation is set
to "On":
Backlight
(regulated)
When backlight
0 ... 3360
regulation is set
to "Off":
Backlight
(controlled)
On
Backlight regu-
lation
Off
Default: On
ADC calibration
Execute
Backlight mar-
Information
gin
6.3 Description of OSD menu
Description
Specified as info in °C
Facility for reading out the current
tab value of the backlight
Backlight sensor regulation
Minimum and maximum regu-
lation
Set brightness of monitor back-
light
Adjustment of overall brightness
to ambient lighting.
Backlight inverter value
The backlight regulation (backlight
sensor) is switched on or off
Automatically calibrate A/D
converter for the applied video
level
The video level range of the sys-
tem is checked, and the monitor
set accordingly. This results in op-
timum adjustment of the individual
RGB A/D converters to the video
source.
The calibration results in a larger
and more flexible video level
range (e.g., the full brightness is
also achieved at 700 mV if the
video level is limited at this value).
Note: A specific test pattern and
timing are prerequisites!
The A/D converters have already
been factory-set and need not be
readjusted.
Remaining margin of backlight
that can be used for regulation
Note: The data is only displayed
when the panel has warmed up
(after approx. 20 min). The dis-
played value is for information
only.
Start­up
41

Advertisement

Table of Contents
loading

This manual is also suitable for:

Smd 19102

Table of Contents