Eizo SCD 19102 Instructions For Use Manual page 26

1.3 mp 19" lcd monitor
Hide thumbs Also See for SCD 19102:
Table of Contents

Advertisement

Connecting
5.9 Serial interface
* Photometer is a generic term for the measuring instrument suggested by EIZO GmbH Dis-
play Technologies:
Advanced Serial Luminance Meter with adapter from 9-pin Sub-D to 6-pin mini-DIN
socket **.
** Service mode! Not permitted in the patient environment.
If the monitor connection is made via the serial bus (see Fig. "Possible configuration – serial
bus mode"), the setting "Serial bus" must first have been activated using "On Screen Dis-
play → Servicelevel 2 → Others". In this case (also under "On Screen Display → Servicelevel
2 → Others") the bus address has to be assigned for each device (1, 2, 3, etc. - assign differ-
ent numbers for the monitors connected via the same bus connection).
Settings, measurements, calibration, and automated quality tests can be performed using
QA software.
See also
Connection panel [} 19]
Menu functions [} 29]
26
SCD 19102 / SMD 19102
Instructions for Use, 08/2015