Eizo SCD 19102 Instructions For Use Manual page 59

1.3 mp 19" lcd monitor
Hide thumbs Also See for SCD 19102:
Table of Contents

Advertisement

Known data
If some of the timing data are already known, for example the resolution, they should first be
entered in
"Blind mode
If the resolution is known, it can be used for the approximation of "Horizontal lines (total)".
The following applies: "Horizontal lines (total)" > "Horizontal resolution". If, when setting
"Horizontal lines (total)", the "Horizontal resolution" is larger, the latter must be reduced. A
larger "Horizontal resolution" can result in the image being split vertically.
Sequence for setting a timing
1. Determine start values using auto function
2. Optimize scanning frequency
3. Optimize geometry and resolution
Step-by-step instructions for setting a timing can be found in the next three sections.
Determine start values using auto function
The following commands can be used to trigger automatic determination of the timing data,
and provides rough approximation of the applied timing.
1. Open a test pattern with a clearly defined border (e.g. SMPTE image).
2. Open the "Live mode" menu: Main menu → Servicelevel 2 → Force Mode → Live mode.
3. Select either "For static pictures" or "For motion pictures" in the "Live mode" menu and
confirm using key 2.
SCD 19102 / SMD 19102
Instructions for Use, 08/2015
[} 56]" in order to simplify the next steps.
Start­up
6.4 System settings