Customer-Specific Visual Inspection; Customer-Specific Test Procedures; Customer-Specific Device Templates - Benning ST 755+ Operating Manual

Table of Contents

Advertisement

Functions
7.4 Expert settings
Saving limits
The modified limits are stored temporarily (until the next restart of the device). The standard test
procedures in the automatic measuring mode will not be changed. Newly created customer-
specific test procedures receive the standard limits from the device, but not the temporarily
changed limits.
Resetting limits
All temporarily changed customer-specific limits are reset to the default limits (factory settings).
Overwriting limits
All changed limits are set permanently as new default limits. The change applies after
confirmation of the "Overwrite" button and affects all newly created customer-specific test
procedures and all manual tests. The limits of the standard test procedures will not be changed.
Automatic determination of limits
If test values are calculated from different limits, they are automatically determined by the
device.
Example for R
• The preset and admissible resistance value for a line length of 5 m and a cross-section of
up to 1.5 mm
• For each additional line length of 7.5 m, the admissible resistance value increases by 0.1 Ω
each.
• The maximum admissible resistance value is 1.0 Ω.
• If you increase the line length to a value from >5 m to 12.5 m, the device automatically
calculates the admissible resistance value.
• For cross-sections of more than 1.5 mm
using the following equation: R = ρ * ℓ/Α + 0.1 Ω
7.4.2

Customer-specific visual inspection

The "Customer-specific visual inspection" [} page 61] menu is intended to manage individual
test questions.
Here, you can create and save test questions. The saved test questions will be displayed on the
"Questions" screen.
7.4.3

Customer-specific test procedures

You can access the created customer-specific test procedures via the automatic and manual
tests of the device. Customer-specific test procedures can be assigned to both newly created
test samples and already existing test samples [} page 61].
7.4.4

Customer-specific device templates

The "Customer-specific device templates" [} page 63] menu is intended to manage test
sample templates.
5284 / 08/2023 en
acc. to VDE 0701 / 0702:
PE
2
is 0.3 Ω.
BENNING ST 755+ / ST 760+
2
and other cable lengths, the limit is calculated
37