Table 11 Test View Buttons; Table 12 "Test Result" Screen - Benning ST 755+ Operating Manual

Table of Contents

Advertisement

Button
Cable (On/Off)
Offset (On/Off)
Done
Table 11: Test view buttons
Eliminating measuring errors
If a measured value is outside the set limits, it will be displayed in red font. In this case, the test
sample will not pass the test.
1. Establish a proper connection between the test probe and the contact point of the test
sample.
2. Tap the "Reset" button while the measurement is active or running.
The maximum value will be overwritten by the currently measured value.
If the measured value is lower than the defined limit, the test is passed.
Test report "Test result"
At the end of a test procedure, the test result is displayed in the form of a test report. Fill in the
following information:
Setting options
Device
*Manual test
Test type
Note
Table 12: "Test result" screen
32
Description
For reversing the polarity of the test voltage
during manual individual tests.
Is enabled for the detection of defects on
cables / mains connection cables.
ON = The protective conductor is continuously
checked for cable breaks. For this test step,
the "RPE test repetition" [} page 56] setting is
temporarily switched off.
OFF = The continuous check for cable breaks
of the protective conductor is switched off. For
this test step, the "RPE test repetition" setting
remains switched on.
Enables / disables the offset by means of a
measuring adapter during the test. To do this,
specify the internal resistance of the measuring
adapter under: "Settings >System
settings >RPE resist. test adapter"
Completes the test and goes to the "Test
result" test report.
Description
Select the test sample:
Tap the right side of the "Change" line to select the test sample from
the database or to create a new test sample.
Select the test type:
• Periodic testing (VDE 0702)
• Test after modification (VDE 0701)
• Test after repair (VDE 0701)
• Receiving inspection (VDE 0702)
Add a comment regarding the test / test sample:
Tap the right side of the "Change" line to open an input window for
adding comments.
BENNING ST 755+ / ST 760+
General operation
6.6 Operating the test view