Alarms; Shut-Down - Dräger Babytherm 8000 OC Instructions For Use Manual

Resuscitation/open care unit
Table of Contents

Advertisement

Alarms

Deviation from specified temperature
In case of deviations greater than ±1 °C between the
specified and actual mattress temperature:
1 The red LED ±1 °C flashes and an intermittent tone
sounds, and
2 the display flashes.
The intermittent tone can be switched off for
approximately 15 minutes:
3 Press the G button, its yellow LED is lit.
1 The red LED is lit continuously.
When the actual mattress temperature has returned to a
value within ±1 °C of the specified temperature:
1 The red LED ±1 °C turns off, and the intermittent
tone goes off.
After the mattress heating is switched on:
– The alarm tone is suppressed for 1 hour (warm-up).
1 The red LED ±1 °C is lit.
3 The yellow LED in the G button is lit.
40 °C alarm
4 The red LED 40 °C flashes, and a continuous tone
sounds.
5 The display flashes.
The continuous tone cannot be suppressed.
The alarm stops automatically when the temperature
decreases.

Shut-down

Only for processing when changing patients.
Press the power switch all the way = OFF.
Then:
Disconnect from mains power supply.
Do this in the specified sequence. Otherwise, the
mains failure alarm will be activated.