Recorder Check; Setting Time And Date - ZOLL PD 1400 Operator's Manual

Pacemaker/defibrillator
Hide thumbs Also See for PD 1400:
Table of Contents

Advertisement

Checks and Procedures
If "TEST FAILED" appears, contact appropriate technical personnel or ZOLL
immediately.
RECORDER CHECK
Press the RECORDER ON-OFF button. The recorder will run until you press the
button again.
While the recorder is running, press and hold the up and down arrows located next
to the alarm set button. This will generate calibration pulses.
lnspect the recorder waveform for uniformity and darkness.
lnspect for uniformity of annotation characters and completeness of words.
Check recorder speed by verifying that a new calibration pulse appears
approximately every 1 3 small divisions (1.3 cm) and is 10+l m m high.
I
Check for adequate supply of paper.
CHANGING PAPER
Press the recorder release button. The door and
paper carriage will tilt up.
Remove the empty or low paper roll.
Place a new roll of thermal paper between the
two side knobs with the paper coming off the
top of the roll and the grid facing down.
Pull enough paper off the roll so that the paper
extends out of the recorder when the recorder
door is closed
Close recorder door.
SETTING TIME
Check the time and date on the recorder annotation. If it is not correct, set as
AND DATE
follows:
Turn the selector switch to OFF.