Room Layouts; Room Dimension Requirements; Room Layout Drawings - GE Innova IGS 6 Preinstallation Manual

Table of Contents

Advertisement

Equipment Requirements
System Compatibilities Cross-Reference – 19" LCD Monitors Sub-System
Product Name
LCD 19" EIZO MX193 Color Monitor
Eizo 19" LCD HB color monitor
RX150 GE

2.2 Room Layouts

2.2.1 Room Dimension Requirements

For Room size dimensions, refer to
For additional details, refer to

2.2.2 Room Layout Drawings

2.2.2.1 Patient Room Layout
WARNING
LOCATION IN TECHNICAL ROOM FOR ELECTRICAL CABINETS IS MANDATORY. THE
ELECTRONIC CABINETS (C1, C2, OPTIONAL LD CABINET, COOLIX 4100 CHILLER,
DETECTOR CONDITIONER AND FLUORO UPS WHEN INSTALLED ) INCLUDE FANS
THAT ARE CREATING AIR-CIRCULATION OF PULSED-AIR. WHEN THIS PULSED AIR
IS IN AN ENVIRONMENT THAT MAY CONTAIN AIRBORNE PATHOGENS LIKE AN
EXAM ROOM/CONTROL ROOM, THERE IS A RISK OF TRANSMISSION OF THESE
AIRBORNE PATHOGENS FROM PATIENTS TO OTHER PATIENTS OR CLINICAL
PERSONNEL (NOSOCOMIAL DISEASES).
TO REDUCE THIS RISK, THE ELECTRONIC CABINETS MUST BE INSTALLED IN A
ROOM SEPARATED FROM EXAM ROOM/CONTROL ROOM, I.E., TECHNICAL ROOM.
Table 2-8 Exam room dimensions
Room Dimensions
Recommended:
Minimum:
72
Pre-Installation Document Num-
sm 5499528-1-8EN
2.2.2.1 Patient Room Layout on page
2.2.3 Room Layout Considerations on page
Length x Width
11570 mm x 8000 mm (38 ft 0 in x 26 ft 2
in)
6900 mm x 4400 mm (22 ft 8 in x 14 ft 5 in)
TM
Innova
IGS 6
ber
provided by EIZO
Ceiling Height
2845 mm ± 5 mm (9 ft 4 in ± 0.2 in) is man-
datory ceiling height
Catalog Number
72.
80.
5750182-1EN 3