Injector Mounting Requirements - GE Optima IGS 320 Preinstallation Manual

Table of Contents

Advertisement

3.3.4 Injector Mounting Requirements

130
CAUTION
Table accessory rail load consideration:
The maximum load per table accessory rail is 40 kg (88 lbs) at 150 mm (0.49 ft)
(60 N.m or 44.25 ft/lbs). Therefore:
Only light extra load not exceeding 5 kg (11 lbs) at 100 mm (0.33 ft) (i.e IV
pole with its accessories, pressure head...) is authorized on the same table
accessory rail as the injector.
Never install injector and radiation protection on the same table accessory
rail.
Typical installation on the front table accessory rail is Smart handle or Smart
box, Table Side System Control (TSSC), Table panning device and cables
support.
If needed an optional rail can be installed at table foot end of the Omega V
table for other options.
Optima IGS 320, Optima IGS 330 Pre-Installation Manual
Direction 5537562-1-1EN, Revision 3
3 Room Structural Requirements