33. Send the MXG a "LAN1" peer to peer message:
Using Interactive LXI or the Keysight IO Libraries TMFramework LXI library: send a "LAN1"
peer to peer message to the MXG.
This will start the synchronization sequence.
34. The MXA waits for the "OperationComplete" instrument event:
Detect the peer to peer traffic using Interactive LXI. To programmatically listen for LXI peer to
peer messages, use the Keysight IO Libraries TMFramework LXI library.
When the MXA has completed its list, it will send the peer to peer message
"OperationComplete" with a rising edge to MXG.
35. Send query to MXA for the measurement results:
:FETCh:LIST?
Figure 2-9 Illustration of a List Sweep Measurement Between an MXA and an MXG using Peer to Peer Messages.